Quantifying the impact of coronary-related experimental simplifications on flow features in native and neo-sinus following transcatheter aortic valve implantation.
Yuhao Wei, Fei Chen, Tianai Wang, Sebastian V Jansen, Tinghui Zheng, Mao Chen, Peng Wu
Department of Mechanics & Engineering, College of Architecture & Environment, Sichuan University, Chengdu, 610065, China. · Department of Cardiology, West China Hospital, Sichuan University, No. 37 Guo Xue Xiang, Chengdu, 610041, Sichuan Province, China.
Key points
From the paper's conclusion- Under the present conditions, coronary flow substantially increased velocity magnitude, vorticity, and particle washout within the corresponding native sinus and neo-sinus after self-expanding valve implantation.
Abstract
Existing in vitro and numerical studies lack consensus regarding whether and how coronary arteries should be incorporated. This study aims to systematically investigate the effects of coronary artery outlets on the hemodynamic environment within the native sinus and neo-sinus after transcatheter aortic valve implantation (TAVI). Three idealized aortic root models (without coronaries, single coronary, and bilateral coronaries) were fabricated. A VENUS self-expanding valve was implanted at five depths (0 mm, ± 5 mm and ± 10 mm). A pulsatile in vitro flow platform combined with particle image velocimetry (PIV) was applied to quantify velocity fields, vorticity, and particle washout. Correlations between implantation depth and hemodynamic parameters were further assessed. In control models, mean native sinus velocity without coronaries was 0.58 ± 0.49 cm/s and decreased further after TAVI. Introducing a single coronary increased mean velocity to 1.34 ± 0.95 cm/s and generated high-velocity jets (> 10 cm/s) near the ostium; bilateral coronaries produced comparable effects. Vorticity decreased in all post-TAVI configurations. Particle washout analysis demonstrated pronounced stasis without coronary flow but markedly improved clearance when coronary inflow was present. With coronary flow, particle washout was markedly enhanced compared with the no-coronary condition, but did not vary monotonically with implantation depth; instead, it appeared to be governed by the combined effects of local flow environment. Under the present conditions, coronary flow substantially increased velocity magnitude, vorticity, and particle washout within the corresponding native sinus and neo-sinus after self-expanding valve implantation. Neglecting coronary outlets may lead to a substantial underestimation of sinus flow velocity and washout. However, when evaluating the hemodynamics of an individual coronary sinus, inclusion of its corresponding coronary artery alone is likely to be sufficient to capture the essential flow characteristics.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Transcatheter aortic valve canting creates cusp-specific thrombogenic microenvironments through altered sinus hemodynamics.JTCVS structural and endovascular · 2026
- Computational fluid dynamics assessment of hemodynamic effects of catheter navigation during transcatheter aortic valve implantation.The International journal of artificial organs · 2026
- Association of Implantation Depth With Hemodynamics and Mechanical Consequences in Small Valve-In-Valve TAVR.JACC. Asia · 2026
- Experimental Bench Validation of a Computational Fluid-Structure Interaction Model of Transcatheter Aortic Valve DynamicsarXiv · 2026
- Influence of Cardiac Phase on Valve-to-Coronary Ostia Distance in Valve-in-Valve Transcatheter Aortic Valve Replacement.Canadian Association of Radiologists journal = Journal l'Association canadienne des radiologistes · 2026
- Computational modeling of thrombosis after transcatheter aortic valve implantation: A systematic review of hemodynamic and mechanical risk metrics.Computer methods and programs in biomedicine · 2026