Phase-dependent agreement between noninvasive and invasive blood pressure monitoring during transcatheter aortic valve replacement: a prospective observational study.
Raffael Pereira Cezar Zamper, Ahmed Abdelaatti, Ricardo Hideo Tachibana, Luiz Guilherme Villares da Costa, Michael W A Chu, Rodrigo Bagur, Philip M Jones
Department of Anesthesia and Perioperative Medicine, Schulich School of Medicine and Dentistry, Western University, London, ON, Canada. rzamper@uwo.ca. · Department of Anesthesia and Perioperative Medicine, Schulich School of Medicine and Dentistry, Western University, London, ON, Canada.
Key points
From the paper's conclusion- ClearSight showed phase-dependent agreement with IAP during TAVR, typically underestimating systolic and mean pressures, with improved agreement during hemodynamic stability, most notably after protamine. These findings support ClearSight as an adjunct rather than a replacement for invasive arterial monitoring during TAVR. Broader validation studies are warranted.
Abstract
PURPOSE: We aimed to evaluate the phase-dependent agreement between continuous noninvasive blood pressure monitoring using ClearSight™ (Edwards Lifesciences Corp., Irvine, CA, USA) and invasive arterial pressure (IAP) across distinct intraoperative phases of transcatheter aortic valve replacement (TAVR). METHODS: We conducted a single-centre observational study at a tertiary academic hospital performing TAVR. We included 21 patients undergoing TAVR, contributing 987 paired invasive and noninvasive blood pressure measurements. We recorded noninvasive blood pressure monitoring using ClearSight concurrently with standard IAP monitoring as part of routine clinical care. We analyzed repeated paired measurements using Bland-Altman methods extended with linear mixed-effects modeling, with additional sensitivity analyses performed to assess robustness. RESULTS: We compared invasive and ClearSight-derived systolic, mean, and diastolic pressures. ClearSight underestimated systolic and mean pressures and overestimated diastolic pressure. After excluding outliers identified by quantile-quantile plot diagnostics, mean biases (IAP minus ClearSight) were +14 mm Hg (systolic), +6 mm Hg (mean), and -3 mm Hg (diastolic). Limits of agreement narrowed after outlier removal, and intraclass correlation coefficients improved (systolic, 0.69 to 0.72; mean, 0.65 to 0.68; diastolic, 0.59 to 0.63). Phase-specific analysis showed the widest disagreement during induction/stabilization, improved accuracy during valve deployment, and the best agreement, particularly for mean arterial pressure, after protamine administration, when vascular tone and hemodynamic stability were restored. CONCLUSIONS: ClearSight showed phase-dependent agreement with IAP during TAVR, typically underestimating systolic and mean pressures, with improved agreement during hemodynamic stability, most notably after protamine. These findings support ClearSight as an adjunct rather than a replacement for invasive arterial monitoring during TAVR. Broader validation studies are warranted.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Accuracy of cuffless blood pressure devices compared with invasive arterial line monitoring in high-acuity settings: a random-effects meta-analysisEuropean Heart Journal · 2026
- Interchangeability Between Pleth Variability Index and Internal Jugular Vein Variability Index During Laparoscopic Living Donor Nephrectomy: A Prospective Observational Concordance StudymedRxiv · 2026
- Comparing noninvasive blood pressure monitoring technologies: the challenge of between-arm differences.British journal of anaesthesia · 2026
- Real-world performance of a ring-type cuffless blood pressure device compared with ambulatory blood pressure monitoring.Clinical hypertension · 2026
- Long-Term Stability and Blood Pressure Change Tracking with a Cuffless Wearable Device in Hypertensive Patients.Blood pressure · 2026
- Real-Time Hemodynamic Guidance During Valve-in-Valve TAVI.JACC. Case reports · 2026