Impact of dynamic pump speed adjustment versus constant speed in left ventricle assist device patients on exercise performance: A systematic review and meta-analysis.
Sarah Quick Lourenço de Lima, Vitória Freitas Silva, Matheus de Medeiros Fernandes, Rafaela Pereira Silva, João Filipe Cavalcante Uchoa Furtado, Lívia Liberata Barbosa Bandeira, Juliana Giorgi, Rodolfo Alves Lopes
Department of Medicine, Federal University of Juiz de Fora, Juiz de Fora, Minas Gerais, Brazil. · Department of Cardiology, Heart Institute (Incor), University of São Paulo Medical School, São Paulo, São Paulo, Brazil.
Key points
From the paper's conclusion- No differences were observed in peak cardiac output (MD 0.12 L/min; 95% CI -0.74 to 0.98; p = 0.786; I2 = 67%) or ventilatory efficiency, assessed by the VE/VCO2 slope (MD -1.58; 95% CI -3.25 to 0.09; p = 0.063; I2 = 76%).ConclusionDynamic LVAD speed adjustment may improve exercise capacity, as suggested by higher peak VO2, although peak cardiac output and ventilatory efficiency showed no clear im
Abstract
BackgroundAdjusting pump speed in patients with left ventricular assist devices (LVADs) during exercise may improve hemodynamics and enhance functional capacity in advanced heart failure, but its clinical benefit over fixed-speed operation remains uncertain. We evaluated whether dynamic, exercise-driven LVAD speed modulation improves physical performance compared to fixed-speed settings.Methods and resultsPubMed, Embase, and the Cochrane Library were systematically searched for studies comparing dynamic adjustment versus fixed pump speed control during exercise in patients with LVADs. Continuous outcomes were analyzed using mean differences (MD), and binary outcomes using risk ratios (RR), with 95% confidence intervals (CIs). Statistical analysis was performed using R (version 4.3.1). Eight studies were included, with study sample sizes totaling 124 participants; seven used crossover designs and one used a non-randomized repeated-measures design. Potential overlap between two reports from Jung et al. precluded interpretation of this total as the number of unique participants. Dynamic pump-speed modulation was associated with higher peak oxygen uptake (VO2) (MD 1.07 mL/kg/min; 95% CI 0.73 to 1.41; p < 0.001; I2 = 42%) and LVAD pump flow at peak/end of exercise (MD 1.24 L/min; 95% CI 0.68 to 1.79; p < 0.001; I2 = 89%). No differences were observed in peak cardiac output (MD 0.12 L/min; 95% CI -0.74 to 0.98; p = 0.786; I2 = 67%) or ventilatory efficiency, assessed by the VE/VCO2 slope (MD -1.58; 95% CI -3.25 to 0.09; p = 0.063; I2 = 76%).ConclusionDynamic LVAD speed adjustment may improve exercise capacity, as suggested by higher peak VO2, although peak cardiac output and ventilatory efficiency showed no clear improvement. Given the small studies, heterogeneous protocols and possible participant overlap, it remains uncertain whether this apparent benefit is sustained or clinically meaningful to patients.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Ventriculo-Arterial Coupling and Speed Modulations in Rotodynamic Blood Pumps.American journal of physiology. Heart and circulatory physiology · 2026
- Restoring Arterial Pulsatility via Speed Modulation in Ventricular Assist Devices: A Structured Narrative Review and Mechanistic Windkessel-Based Analysis.Artificial organs · 2026
- Patient-centred medical tools for sustained motivation in cardiac rehabilitation of patients with heart failure: a multicentre randomized controlled trialEuropean Heart Journal · 2026
- In Vivo Comparison of Algorithms Estimating Left Ventricular Stroke Work From LVAD-Derived Pump Parameters.Artificial organs · 2026
- Efficacy and safety of exercise rehabilitation across the valve intervention pathway: A systematic review and meta-analysis.Frontiers in cardiovascular medicine · 2026