Vector Flow Mapping Energy Loss and Pump Flow in Left Ventricular Assist Device Patients: A Transesophageal Echocardiography Pilot Study.
Hiroya Tsujimoto, Hiroko Kanazawa, Akito Tsukinaga, Akira Shimokawa, Masahiro Morinaga, Yoshiaki Ito, Shinnosuke Miura, Kenji Yoshitani,
From the Department of Anesthesiology and Pain Relief Center, University of Tokyo Hospital, Bunkyo-ku, Tokyo, Japan. · Department of Anesthesiology, National Cerebral and Cardiovascular Center, Suita, Osaka, Japan.
Key points
From the paper's conclusion- We examined whether vector flow mapping-derived energy loss (EL), a measure of viscous dissipation related to spatial velocity gradients, was associated with controller-displayed pump flow.
Abstract
Left ventricular assist device (LVAD) support redirects intraventricular flow toward the inflow cannula. We examined whether vector flow mapping-derived energy loss (EL), a measure of viscous dissipation related to spatial velocity gradients, was associated with controller-displayed pump flow. This prospective single-center pilot enrolled 33 patients undergoing LVAD-related surgery with intraoperative transesophageal echocardiography; 31 HeartMate 3 recipients were analyzable. The maximum available peak EL across three views was the primary measure. Log-transformed maximum peak EL was associated with displayed flow after adjustment for pump speed and left ventricular end-diastolic diameter (B = 0.181, 95% confidence interval [CI]: 0.033-0.330), cannula angle (B = 0.204, 95% CI: 0.052-0.356), or aortic regurgitation grade (B = 0.204, 95% CI: 0.050-0.358). Among 28 complete cases, associations persisted using the mean of three view-specific peaks (B = 0.213-0.245; p = 0.007-0.019). In a 10-case interobserver reanalysis, maximum peak EL had an intraclass correlation coefficient of 0.759 (95% CI: 0.294-0.934), with ratio limits of agreement of 0.56-2.41. Cannula angle and aortic regurgitation grade were not associated with EL. These preliminary findings support larger prospective studies but do not establish mechanism, measurement interchangeability, or clinical utility.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Pulmonary arterial elastance after LVAD implantation as a predictor of late right ventricular failure beyond six months of supportEuropean Heart Journal · 2026
- Biventricular Pressure-Volume Loop Analysis in an LVAD Patient Optimized With Impella RP Flex.Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2026
- Tricuspid annular plane systolic excursion/pulmonary vascular resistance index predicts right ventricular assist device requirement and mortality after magnetically levitated centrifugal-flow left ventricular assist device implantation.JTCVS open · 2026
- Pre-implant right heart catheterization and liver biomarkers in left ventricular assist device candidatesEuropean Heart Journal · 2026
- A novel magnetically levitated intrapericardial left ventricular assist system in advanced heart failure—2-year results from a prospective, multi-centre studyEuropean Journal of Cardio-Thoracic Surgery · 2026
- Towards Modeling the Hemodynamic Impact of Mitral and Aortic Valve Repair in Patients with Left Ventricular Assist Devices.ArXiv · 2026