Under pressure- test, test, maybe? A survey of heat exchanger integrity testing at cardiac centres in Australia and New Zealand.
Martin Gill
Perfusion Department, John Hunter Hospital, Newcastle, NSW, Australia.
Key points
From the paper's conclusion- This study aimed to identify current practices across Australia and New Zealand and assess the mechanical efficacy of manufacturers recommended testing protocols.MethodsA telephone survey was conducted of the most senior perfusionist available at all 80 cardiac surgical hospitals in Australia and New Zealand regarding HE testing customs and known failures.
Abstract
IntroductionHeat exchanger (HE) integrity is vital for patient safety during cardiopulmonary bypass (CPB), as failures can lead to severe complications including haemolysis and toxemia. While manufacturers mandate pre-use integrity testing, adherence and methods vary significantly. This study aimed to identify current practices across Australia and New Zealand and assess the mechanical efficacy of manufacturers recommended testing protocols.MethodsA telephone survey was conducted of the most senior perfusionist available at all 80 cardiac surgical hospitals in Australia and New Zealand regarding HE testing customs and known failures. Simultaneously, an in vitro assessment measured the pressures generated by two common heater cooler units (HCUs)- the 3T and HCU40, in conjunction with oxygenator and cardioplegia HEs to evaluate the generated pressures during usage.ResultsNine unverified historical HE failure events were recalled by surveyed senior perfusionists across the region. While 100% of programs were successfully surveyed, only 49% complete perfusion teams consistently perform pre-clinical integrity testing. Of those that test, the most common method (82%) is running water through the HE, as per manufacturers instructions. However, in vitro data revealed that while the HCU40 maintains positive pressure, the 3T HCU generates negative pressure at the HE outlet during use. This negative pressure gradient creates a theoretical risk of masking defects near the outlet during a standard 'water-run' test, as a positive pressure gradient is not uniformly maintained across the HE. Such an occurrence would render the manufacturers "water-run" test insufficient for detecting defects near the outlet.ConclusionCurrent HE testing practices are non-uniform and often limited by logistical barriers. Reliance on fluid-based testing is hemodynamically flawed in certain HCU configurations. To enhance adherence and safety, manufacturers should validate gas pressure testing for end users, and perfusion departments should adopt standardised, positive-pressure testing protocols and documented emergency failure algorithms.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Complete ventilation failure due to a defective heat and moisture exchange filter.Anaesthesia reports · 2026
- Oxygenator deterioration in a standardized in vitro setup: pressure drop, CT-based clot burden, and hemolysisCurrent Directions in Biomedical Engineering · 2026
- Ex Vivo Heart Perfusion Versus Cold Storage in Prolonged Preservation Period: A Juvenile Porcine Experimental Model.ASAIO journal (American Society for Artificial Internal Organs : 1992) · 2026
- The Hammer Effect: Variables that Promote Air Migration Across Semi-Permeable Membranes During Cardiopulmonary BypassThe Journal of ExtraCorporeal Technology · 2026
- Phantom system for endotracheal tube cuff seal testing during mechanical ventilationMedical & Biological Engineering & Computing · 2026
- Practice Prevalence of Blood Transfusion Practices in the Tertiary Cardiac Center of Nepal.Kathmandu University medical journal (KUMJ) · 2026