Complete ventilation failure due to a defective heat and moisture exchange filter.
Kathryn Taylor, Yvonne Csanyi-Fritz
Department of Anaesthesia BC Children's Hospital Vancouver Canada.
Key points
From the paper's conclusion- We report a case of a 3‐year‐old patient where a complete circuit obstruction due to a defective heat and moisture exchange filter occurred.
Abstract
Heat and moisture exchange filters are routinely used in anaesthesia to provide passive humidification and protect the breathing circuit. We report a case of a 3-year-old patient where a complete circuit obstruction due to a defective heat and moisture exchange filter occurred. This resulted in a total inability to ventilate the patient's lungs, initially attributed to bronchospasm, followed by the need for cardiopulmonary resuscitation. Manufacturing defects resulting in such critical incidents are rare but remain an important potential differential when managing difficulty in ventilation. This incident underscores the need to maintain a high index of suspicion for equipment malfunction when difficulty in ventilation arises, even in the presence of plausible clinical causes. Automated anaesthetic machine checks do not assess circuit patency; therefore, manual verification, such as the two-bag test, remains essential following any circuit change. This case also illustrates how human factors and system gaps can align to permit equipment defects to reach the patient, while effective teamwork and crisis management can help to mitigate harm. Routine inclusion of circuit patency testing, simulation-based training and rigorous manufacturing quality control are suggested for preventing recurrence and enhancing patient safety in anaesthetic practice.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Misconnection of Oxygen Tubing to the "Smart Cuff Manager" Causing Cuff Explosion: A Preventable Equipment Error.A&A practice · 2026
- Under pressure- test, test, maybe? A survey of heat exchanger integrity testing at cardiac centres in Australia and New Zealand.Perfusion · 2026
- Moisture ingress in a hospital medical air system secondary to dryer failure leading to service disruption.Anaesthesia reports · 2026
- Airway Obstruction With Severe Hypoxemia Caused by Malformation of an Endotracheal Tube Cuff: A Case Report.Clinical case reports · 2026
- Phantom system for endotracheal tube cuff seal testing during mechanical ventilationMedical & Biological Engineering & Computing · 2026
- Inadvertent Epidural Administration of Diluted Unfractionated Heparin During Combined Spinal-Epidural Anesthesia: A Case Report and Patient Safety Analysis.Local and regional anesthesia · 2026