Circumferential Identification of Peripheral Anterior Synechiae by Air-Puff Dynamic Anterior Segment Optical Coherence Tomography.
Xikai Wei, Cheng Lu, Jianli Ren, Kaixi Zheng, Shaodan Zhang, Peijuan Wang, Dexi Zhu
Eye Hospital and School of Ophthalmology and Optometry, Wenzhou Medical University, Wenzhou, Zhejiang, People's Republic of China.
Key points
From the paper's conclusion- The air-puff DAS-OCT system enables the noninvasive, dynamic, and circumferential identification of PAS, demonstrating high diagnostic performance.
Abstract
PURPOSE: The purpose of this study was to objectively and accurately identify peripheral anterior synechiae (PAS) circumferentially in eyes with angle-closure glaucoma (ACG). METHODS: Anterior chamber angles (ACAs) were classified as PAS or non-PAS (NPAS) groups by indentation gonioscopy. A custom-built air-puff dynamic anterior segment optical coherence tomography (DAS-OCT) system with an innovative 4-leaf clover-scan pattern acquired 10 frames per ACA along 8 radial directions at 45-degree intervals within 65 ms. Dynamic parameters including the differences in angle opening distance at 750 µm from the scleral spur (AOD750) and trabecular-iris space area at 750 µm anterior to the scleral spur (TISA750) before and after air-puff stimulation, were extracted to establish a binary logistic regression model. Model performance was evaluated in the testing set. RESULTS: Analysis of 68 images from 9 healthy eyes confirmed high measurement repeatability (intraclass correlation coefficient [ICC] = 0.821 for AOD750dif; 0.775 for TISA750dif). A total of 143 anterior segment OCT (AS-OCT) images from 20 ACG eyes were analyzed, with 100 assigned to the training set and 43 to the testing set. Significant differences in AOD750dif and TISA750dif were identified between the NPAS and PAS groups (P < 0.001). The dual-parameter model achieved an area under the curve (AUC) of 0.924 in the training set and 0.913 in the testing set, with an overall accuracy of 83.7%. CONCLUSIONS: The air-puff DAS-OCT system enables the noninvasive, dynamic, and circumferential identification of PAS, demonstrating high diagnostic performance. TRANSLATIONAL RELEVANCE: This non-contact imaging technology offers a quantitative alternative to indentation gonioscopy, potentially improving ACG screening efficiency and guiding clinical management.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Quantitative Assessment of Anterior Segment OCT Parameters and Their Association with Intraocular Pressure in Primary Angle-Closure Disease.Diagnostics (Basel, Switzerland) · 2026
- Diagnostic ability of Cirrus anterior segment optical coherence tomography to differentiate between binary grades of narrow angles seen on gonioscopy.Indian journal of ophthalmology · 2026
- Anterior Chamber Angle Alterations in Bilateral Acute Iris Transillumination Syndrome.Ocular immunology and inflammation · 2026
- Correlation between anterior segment optical coherence tomography parameters and visual field results after a single episode of acute angle closure glaucoma.Journal francais d'ophtalmologie · 2026
- The Use of Artificial Intelligence for Estimating Trabecular-Iris Angle from Slit-Lamp Images Developed Using Anterior-Segment Optical Coherence Tomography.Bioengineering (Basel, Switzerland) · 2026
- Anterior segment optical coherence tomography image spatial features in the retro-iris spaceJournal of the European Optical Society-Rapid Publications · 2026