Influence of Screw Access Hole Diameter and Preparation Method on the Fit and Fracture Resistance of Implant-Supported Monolithic Zirconia Crowns.
Min-Gyung Seo, Kyung-Ho Ko, Yoon-Hyuk Huh, Lee-Ra Cho, Chan-Jin Park
Key points
From the paper's conclusion- The 1-mm SAH improved crown fit without reducing fracture resistance, whereas openings of 2 mm or larger reduced fracture resistance. For 3-mm SAHs, preparation method did not affect mean fracture load.
Abstract
PURPOSE: To evaluate how screw access hole (SAH) presence, diameter, and preparation method affect marginal and internal fit and fracture resistance of implant-supported monolithic zirconia crowns. MATERIALS AND METHODS: Seventy-two crowns were assigned to six groups (n=12): no SAH; 1-, 2-, or 3-mm presintered SAHs; a 1-mm SAH enlarged to 3 mm after cementation; or a 3-mm SAH prepared after cementation in crowns without an initial SAH. Marginal and internal fit was assessed in the first four groups by the silicone replica technique. After thermocycling and cyclic loading, fracture load, Weibull parameters, fracture patterns, and scanning electron microscopy findings were evaluated. RESULTS: Overall discrepancy and internal gap were greater without an SAH (184.7 and 182.1 μm) than with an SAH (54.7-65.9 and 60.1-71.1 μm) (P<.05), with no differences among SAH diameters (P>.05). Fracture loads were higher for no SAH and a 1-mm SAH than for all groups with 2- or 3-mm openings (P<.001). Among final 3-mm openings, preparation method did not affect mean fracture load (P>.05), although handpiece-prepared groups had lower Weibull moduli. CONCLUSIONS: The 1-mm SAH improved crown fit without reducing fracture resistance, whereas openings of 2 mm or larger reduced fracture resistance. For 3-mm SAHs, preparation method did not affect mean fracture load.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Effect of metal framework design on the fracture resistance of implant-supported metal-ceramic crowns fabricated by selective laser melting.Journal of prosthodontics : official journal of the American College of Prosthodontists · 2026
- 3D-printing for zirconia abutment manufacturing and screw access hole effect: fatigue behavior of milled zirconia simplified restorations.Odontology · 2026
- Influence of Prosthetic Screw Designs with and Without Titanium Base on the Fracture Resistance of Implant-Supported Zirconia Frameworks.Materials (Basel, Switzerland) · 2026
- Effect of Crown Height Space and Abutment Configuration on the Maximum Recorded Load of Implant-Supported Monolithic Zirconia Crown Assemblies.Materials (Basel, Switzerland) · 2026
- Fit and fracture resistance of 3D-printed single crowns made from highly translucent zirconia.Journal of prosthodontics : official journal of the American College of Prosthodontists · 2026
- Effect of Angulated Screw Channel (ASC) design on the fracture resistance of implant-supported monolithic zirconia and lithium disilicate crowns: an in vitro study.BMC oral health · 2026