Effect of clinical use and autoclave sterilization on the dynamic cyclic fatigue resistance of ProTaper Ultimate instruments.

Taher Al Omari, Aws Khanfur, Kais Bietar, Rashid El Abed

Department of Conservative Dentistry, Faculty of Dentistry, Jordan University of Science and Technology, P.O. Box 3030, Irbid, 22110, Jordan. tmalomari@just.edu.jo. · Department of Conservative Dentistry, Faculty of Dentistry, Jordan University of Science and Technology, P.O. Box 3030, Irbid, 22110, Jordan.

Key points

From the paper's conclusion
  • This study evaluated the effect of clinical use and autoclave sterilization on the dynamic cyclic fatigue resistance of ProTaper Ultimate (PTU) instruments and assessed whether repeated use altered surface roughness of F2 instruments.

Abstract

This study evaluated the effect of clinical use and autoclave sterilization on the dynamic cyclic fatigue resistance of ProTaper Ultimate (PTU) instruments and assessed whether repeated use altered surface roughness of F2 instruments. Thirty-six mandibular molars were instrumented with PTU files under standardized clinical conditions. Files were allocated to 3 groups: new files, files used once and sterilized once, and files used twice, with sterilization after each use. Dynamic cyclic fatigue testing was performed at simulated body temperature (37 °C) using a pecking model. The number of cycles to failure (NCF) was recorded for Slider, Shaper, F1, and F2. Fractured fragment length was measured, fractured surfaces were examined by scanning electron microscopy, and surface roughness was assessed at the coronal, middle, and tip sections of F2 files. The general linear model identified significant effects of instrument type, group, and their interaction on the number of cycles to failure ( P < 0.05). Slider demonstrated a progressive reduction from 3473.17 cycles when unused to 2127.42 cycles after two use-sterilization cycles. File-specific pairwise comparisons identified significant reductions for Slider, whereas the numerical reductions observed for Shaper, F1 and F2 were not statistically significant. Fragment length and F2 surface roughness did not differ significantly among groups (P > 0.05). Clinical use followed by autoclave sterilization was associated with an overall, instrument-dependent reduction in cyclic fatigue resistance. The clearest progressive reduction occurred in Slider, while surface roughness did not differ significantly among F2 groups.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers