Underrecognized Interaction Between PICO Dressings and Pacemakers: A Case From the ICU.
Reginald Caldecott, John Laffey
Department of Anaesthesia and Intensive Care Medicine, University Hospital Galway, Galway, Ireland, hse.ie. · Department of Anaesthesia, School of Medicine, University of Galway, Galway, Ireland, universityofgalway.ie.
Key points
From the paper's conclusion- Magnet-containing medical equipment placed near a cardiac implantable electronic device can trigger magnet-mode pacing. The risk depends on the position of the magnet-bearing component, not on the wound site. An unexplained change in a paced rate should prompt early consideration of a nearby magnetic source, because the remedy may be immediate and at the bedside.
Abstract
BACKGROUND: Modern pacemakers incorporate a magnet-responsive switch that, on exposure to a sufficient static magnetic field, converts the device to fixed-rate asynchronous (magnet-mode) pacing. This is designed behaviour rather than malfunction. Interactions between this feature and magnet-containing medical equipment are underrecognised. The PICO 7 negative pressure wound therapy (NPWT) system contains a magnet in its pump unit, which the manufacturer states can interfere with nearby pacemakers. CASE PRESENTATION: An 81-year-old man was admitted to the High Dependency Unit after right hemicolectomy. He had atrial fibrillation with a slow ventricular response and a permanent pacemaker programmed VVI at 60 beats per minute. Overnight, telemetry showed a sustained paced rate of 100 beats per minute; he was asymptomatic. The PICO 7 pump unit was found resting on his chest, over the pacemaker generator. Repositioning the pump to the foot of the bed returned the paced rate to 60 beats per minute immediately, and device interrogation the following day confirmed normal pacemaker function. DISCUSSION: The fixed rate of 100 beats per minute, its dependence on pump position and immediate resolution on removal are characteristic of magnet-mode activation rather than generic electromagnetic interference. Although the patient was asymptomatic, a sustained rate of 100 beats per minute carried a risk of myocardial ischaemia and heart failure decompensation given his cardiac history. Recognition and a simple bedside intervention resolved the episode. CONCLUSION: Magnet-containing medical equipment placed near a cardiac implantable electronic device can trigger magnet-mode pacing. The risk depends on the position of the magnet-bearing component, not on the wound site. An unexplained change in a paced rate should prompt early consideration of a nearby magnetic source, because the remedy may be immediate and at the bedside.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Sensor-Driven Pacemaker Tachycardia During Chest Physiotherapy: The Hidden Accelerometer Sensor Response in a Leadless Pacemaker in the VDD Mode.Journal of cardiovascular electrophysiology · 2026
- Magnetically gated millimeter-scale capsule power system for leadless electrotherapy.Science advances · 2026
- Temperature Sensor-Mediated Intraoperative Tachycardia in a Leadless Pacemaker During Cardiac Surgery.JACC. Case reports · 2026
- Context-dependent interactions between variable-loop pulsed field ablation and cardiac implantable electronic devices: a controlled preclinical study.European heart journal open · 2026
- A Wearable Morphic Sensor for Unobtrusive Heart Rate and Respiratory Rate Monitoring in ICU Patients: A Pilot Validation Study.Sensors (Basel, Switzerland) · 2026
- Prevalence and Periprocedural Value of Current of Injury at the Implantation of a Passive Fixation Leadless Pacemaker.Pacing and clinical electrophysiology : PACE · 2026