Magnetically gated millimeter-scale capsule power system for leadless electrotherapy.
Feiyao Yang, Yu Tian, Enyi Zhang, Lixue Yang, Ziyue You, Puguang Peng, Han Qian, Yan Du,
Beijing Institute of Nanoenergy and Nanosystems, Chinese Academy of Sciences, Beijing, P. R. China, 101400. · Cardiac Pacing and CIED Center, Beijing Anzhen Hospital. Capital Medical University, Beijing, P. R. China, 100029.
Key points
From the paper's conclusion- Here, these challenges are addressed through a millimeter-scale capsule power system with magnetically gated activation via a reed switch, which eliminates the stringent orientation and implantation-depth requirements of optically controlled approaches.
Abstract
Temporary cardiac pacing is indispensable for bridging transient bradyarrhythmic indications over days to weeks, yet existing systems remain constrained by device dimensions, surgical complexity, patient burden, and intrinsic operational principles. Here, these challenges are addressed through a millimeter-scale capsule power system with magnetically gated activation via a reed switch, which eliminates the stringent orientation and implantation-depth requirements of optically controlled approaches. In contrast to conventional reed switch-based pacemakers, this capsule system extends the functionality of the reed switch by integrating electrode materials directly onto its surface, thereby enabling a monolithic miniaturized architecture. Pulsed magnetic fields wirelessly actuate intermittent switching, enabling controlled discharge, with robust cardiac pacing demonstrated in large animals. A proof-of-concept closed-loop system is further realized, establishing the feasibility of future integration with a wearable electronic module to enable autonomous electrotherapy upon arrhythmia detection. Collectively, this strategy addresses a key limitation in temporary pacing and may provide a versatile framework for the development of temporary electrotherapeutic systems across cardiac, neural, and regenerative applications.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Optogenetic and sonogenetic pacing of unmodified cardiac tissue upon quick integration with specialized biological actuators.bioRxiv : the preprint server for biology · 2026
- A novel septal sensor for continuous cardiac function monitoringEuropean Heart Journal · 2026
- Magnetic field-based actuator-sensor assembly for dynamic mechanical loading and functional analysis of engineered heart tissues in culturebioRxiv · 2026
- Engineering a Bioelectronic Hybrid Heart with Integrated Sensing and Electrical Modulation.Advanced science (Weinheim, Baden-Wurttemberg, Germany) · 2026
- Self-sustaining leadless intracardiac pacemaker powered by triboelectric nanogenerator.Science advances · 2026
- Optical System for Assessment of Atrial and Ventricular Mechanics in Langendorff-perfused Mouse Hearts.Journal of visualized experiments : JoVE · 2026