Pointwise encoding time reduction with radial acquisition (PETRA) to pseudo-CT (computer tomography) mapping for precision focused transcranial ultrasound applications.
Maria Miscouridou, Alisa A Krokhmal, Aaron T Hess, Jon Campbell, Charlotte J Stagg, Eleanor Martin, Bradley E Treeby
Department of Medical Physics and Biomedical Engineering, University College London, London, WC1E 6BT, United Kingdom. · Nuffield Department of Clinical Neurosciences, FMRIB, Wellcome Centre for Integrative Neuroimaging, University of Oxford, Oxford, OX3 9DU, United Kingdom.
Key points
From the paper's conclusion- Here, the pointwise encoding time reduction with radial acquisition (PETRA) sequence was used because of its stronger ability to image bones.
Abstract
Ultrasound simulations used for treatment planning in transcranial ultrasound stimulation (TUS) currently depend on converting computed tomography (CT) images to acoustic properties. Aiming to reduce subject exposure to ionising radiation, recent literature focused on generating pseudo-CT images from different magnetic resonance (MR) imaging sequences. Here, the pointwise encoding time reduction with radial acquisition (PETRA) sequence was used because of its stronger ability to image bones. A dataset of paired CT and PETRA scans was acquired separately for human subjects and ex vivo skulls. Principal component analysis performed on all bone voxels uncovered the affine relationship converting PETRA images to pseudo-CTs. Acoustic simulations based on CTs and pseudo-CTs were performed in k-Plan and k-wave. For in vivo data, the mean absolute errors in focal position, peak pressure, and focal volume were 0.48±0.25 mm, 5.1±4.0% and 5.8±4.4%. The similarity of acoustic fields from ex vivo CTs/pseudo-CTs was compared with experimental data. The corresponding errors were 1.5 ± 1.2 mm, 17.1 ± 14.6% and 32.5 ± 22.5% for CT and 1.6 ± 1.4 mm, 18.2 ± 17.1% and 19.2 ± 15.9% for pseudo-CT. The small errors for in vivo data and the similarity of the experimental errors for both methods validate the use of PETRA-derived pseudo-CT as an alternative to CT, ensuring precise acoustic field predictions. The conversion tool is available at https://github.com/ucl-bug/petra-to-ct.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- MRI-Free CT-to-Ultrasound Skull Registration in Transcranial Focused Ultrasound using CMUT based Integrated Therapeutic and Imaging ArraysarXiv · 2026
- Subwavelength registration of skull-compensating acoustic holograms via nonlinear parametric array feedbackarXiv · 2026
- High‐Resolution 3D UTE MRI for Integrated Radiation‐Free Pulmonary and Osseous Thoracic ImagingNMR in Biomedicine · 2026
- Noninvasive photoacoustic computed mesoscopy for longitudinal brain imagingScience Advances · 2026
- Feasibility of a 2-Minute Multi-Echo UTE Acquisition for Simultaneous CT-Like Bone-Weighted Imaging and Quantitative T2* Mapping of Short-T2 TissuemedRxiv · 2026
- Experimental Validation of Skull Acoustic Modelling Strategies for Transcranial Focused Ultrasound Simulation: A Cross-Comparison Study.Physics in medicine and biology · 2026