Spectral micro-CT for quantitative analysis of calcification in fibrocartilage.
Vittoria Mazzini, Paolo Cardarelli, Andrew Coathup, Eleonora Olivotto, Francesco Grassi, Enrico Tassinari, Simone Velardita, Angelo Taibi,
INFN - Division of Ferrara, Ferrara, Italy. · University of Trieste, Department of Physics, Trieste, Italy.
Key points
From the paper's conclusion- This approach offers a nondestructive and reproducible alternative to histology for assessing calcification in joint tissue, providing quantitative 3D mapping of calcium concentration in intact samples without the need for staining or physical sectioning.
Abstract
PURPOSE: We describe a method for the quantitative analysis of calcification structures in fibrocartilage samples from human hip acetabular labrum, using a lab-based spectral micro-computed tomography ( μ CT ) system. APPROACH: Tissue samples from 26 patients with osteoarthritis and femoroacetabular impingement were analyzed using a spectral μ CT system featuring photon-counting detector with a 62 μ m pixel size operating at two energy thresholds. Material decomposition was applied to the μ CT images to separate calcium and water-like components, resulting in the 3D spatial distribution and quantification of calcium deposits within the tissue. Results were compared with histological analysis using Alizarin Red™ staining. Two calcium crystal deposition scores on a four-grade scale were assigned based on histological and μ CT analysis, respectively. RESULTS: This method provides 3D images with a voxel size of 20 μ m over a sample size up to 3 cm. The minimum detectable calcium concentration is 0.24 g / cm 3 , and calcified structures as small as 49.6 μ m in diameter were identified. The quantification accuracy was validated using a hydroxyapatite calibration phantom. Across the complete sample set, spectral µCT grading demonstrated significant correlation (Spearman correlation coefficient ρ = 0.67 ) and substantial agreement (weighted Cohen's coefficient κ = 0.70 ) with histological assessment. Moreover, the calcification spatial distribution exhibited very good consistency with histological results, supporting the method validity. CONCLUSIONS: This approach offers a nondestructive and reproducible alternative to histology for assessing calcification in joint tissue, providing quantitative 3D mapping of calcium concentration in intact samples without the need for staining or physical sectioning.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Characterizing Meniscal Calcifications with Photon Counting-Based Dual-Energy Computed TomographyAnnals of Biomedical Engineering · 2026
- Added value of material decomposition for measuring brain calcification progression with spectral computed tomography in primary familial brain calcification.European journal of radiology · 2026
- MicroCT-Based Degradation-Induced Sub-Voxel Porosity Mapping: A Case Study of a PHBV/PLGA/TCP Composite Scaffold.Materials (Basel, Switzerland) · 2026
- Site-Specific Material Decomposition Versus Universal Z-Effective: Optimizing Opportunistic Bone Mineral Density Screening via Dual-Energy CT.Journal of imaging · 2026
- Bridging the gap between micro- and nanoscale chemical imaging of biomaterials: a multimodal and multiscale characterization framework.The Analyst · 2026
- A novel external calibration strategy using hollow short-tube hydroxyapatite ceramics for elemental quantitative imaging of human teeth by LA-ICP-MS.Analytical and bioanalytical chemistry · 2026