A novel external calibration strategy using hollow short-tube hydroxyapatite ceramics for elemental quantitative imaging of human teeth by LA-ICP-MS.
Yu Cui, Xin Yang, Dingwen Zhang, Lanlan Jin, Wei Guo, Shenghong Hu
State Key Laboratory of Geomicrobiology and Environmental Changes, School of Earth and Planetary Sciences, China University of Geosciences, Wuhan, 430074, PR China. · State Key Laboratory of Geomicrobiology and Environmental Changes, School of Earth and Planetary Sciences, China University of Geosciences, Wuhan, 430074, PR China. shhu@cug.edu.cn.
Key points
From the paper's conclusion- These results demonstrate that HAP-T exhibits excellent matrix matching with tooth samples in terms of both chemical composition and laser ablation behavior.
Abstract
Teeth provide retrospective information regarding health status, dietary habits, age, and environmental exposures. Laser ablation inductively coupled plasma mass spectrometry (LA-ICP-MS) enables direct solid analysis with high spatial resolution, rendering it well-suited for the microanalysis of teeth. However, accurate quantification is often hampered by the lack of matrix-matched calibration standards. In this work, an external calibration strategy based on hollow short-tube hydroxyapatite ceramics (HAP-T) was established. Due to the one-dimensional tubular architecture, the sintered HAP-T ceramic exhibits high density and mechanical robustness, and its laser ablation behavior was similar to that of tooth samples. The doped trace elements were homogeneously distributed in HAP-T, with relative standard deviations (RSDs) of line scan signals in the range of 7.17-11.07%. The concentrations of doped elements in two synthetic ceramic calibration standards (HAP-T-1 and HAP-T-2) were determined via LA-ICP-MS and pneumatic nebulization ICP-MS (PN-ICP-MS), and excellent consistency was observed between the two analytical techniques. The proposed method was successfully applied to the quantitative elemental imaging of Ca, Mg, Zn, Sr, and Pb in human deciduous tooth sections, achieving high spatial resolution that allows clear visualization of elemental distributions of fine tissue structures including the neonatal line and enamel-dentine junction. These results demonstrate that HAP-T exhibits excellent matrix matching with tooth samples in terms of both chemical composition and laser ablation behavior. This calibration strategy is thus highly promising for the accurate quantitative analysis and elemental imaging of teeth, bones, and potentially other phosphate-based biominerals.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A novel approach for monitoring the spatial distribution and quantitative analysis of micronutrients in plant tissues using laser ablation ICP-MS imaging.Analytical and bioanalytical chemistry · 2026
- A frozen-suspension external calibration strategy for quantitative imaging of Ce in fresh animal soft tissues by CLA-ICP-MS.Talanta · 2026
- Advances in (LA)-ICP-MS techniques and applications for the in elemental analysis the field of cultural heritage: A reviewTalanta · 2026
- Evaluating limits of detection for single-cell elemental bioimaging by LA-ICP-TOFMSTalanta · 2026
- Quantitative LA-ICP-MS Imaging of Elemental Distributions with Protein Correlation in 3D Tumor Models.Analytical chemistry · 2026
- Spectral micro-CT for quantitative analysis of calcification in fibrocartilage.Journal of medical imaging (Bellingham, Wash.) · 2026