Availability of a Digital Refractometer for Measuring Urine Specific Gravity in Dogs.
Tuğrul Kırkulak, Gülten Emek Tuna
Faculty of Veterinary Medicine, Department of Internal Medicine, Tekirdağ Namık Kemal University, Tekirdağ, Turkey. · Faculty of Veterinary Medicine, Department of Internal Medicine, Aydın Adnan Menderes University, Aydın, Turkey.
Key points
From the paper's conclusion- This study demonstrated that USG in dogs can be measured safely, easily, and quickly with the HDR and that the HDR is suitable for clinical use.
Abstract
OBJECTIVE: Urine specific gravity (USG) is used clinically to assess urine concentration and is routinely evaluated with a refractometer. Several studies have compared optical and digital refractometers in assessing USG in dogs. However, these studies have shown conflicting results. Therefore, this study aimed to evaluate the clinical feasibility of a handheld digital refractometer (HDR) (Atago Pal-Usg Dog) for measuring USG. METHODS: Urine samples were collected from 100 dogs with various problems. The results of the HDR and the reference method, the optical handheld refractometer (OHR) (Atago URC-NE), were evaluated using dependent t-tests, Pearson correlation analysis, Passing-Bablok regression analysis, and Bland-Altman tests. Furthermore, the coefficient of variation (CV) and total analytical error (TAE) rates for the HDR were determined. RESULTS: Although a statistically significant difference was determined between the mean USG measured with OHR and HDR, a positive correlation was found between the two refractometers. The Passing-Bablok regression analysis revealed no systematic or proportional bias between the two refractometers. The Bland-Altman test calculated the mean difference between the OHR and HDR measurements as -0.9 units, demonstrating that the USG measured with HDR could be 1.3 units greater or 3 units less than the OHR. When evaluating the CV, the reproducibility of the HDR was quite good, and for the TAE, this value fell within the desired limits for USG. CONCLUSIONS: This study demonstrated that USG in dogs can be measured safely, easily, and quickly with the HDR and that the HDR is suitable for clinical use.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A-376 Analytical Validation of an FDA-Registered Class I Refractometer for Urine Specific Gravity Measurement to Support Regulatory Compliance and Specimen Validity TestingClinical Chemistry · 2026
- A newly derived formula outperforms published formulae for estimating urine osmolality in healthy dogs receiving intravenous fluids.American journal of veterinary research · 2026
- Changing tonometers, changing intraocular pressure? Rethinking intraocular pressure interpretation in dogs.The Journal of small animal practice · 2026
- Diagnostic Evaluation of Automated Urine Particle Digital Imaging Device for Urinary Tract Infection Stewardship: A Cross-Sectional Laboratory-Based Study in Single Indonesian Tertiary Hospital.Research and reports in urology · 2026
- B-303 Evaluating an Automated Urinalysis System: Analytical Performance and Verification StrategiesClinical Chemistry · 2026
- Analytical validation of the Mindray BC-75R Vet hematology analyzer and comparison with the Siemens Advia 2120i for application in dogs and cats.Journal of veterinary diagnostic investigation : official publication of the American Association of Veterinary Laboratory Diagnosticians, Inc · 2026