Subcutaneous Defibrillator for Primary Prevention in Chronic Kidney Disease Stage 3-5: The Single-Arm STEADIED Study.
Tae-Wan Chung, Tae-Hoon Kim, Boyoung Joung, Seung-Jung Park, Eue-Keun Choi, Myung-Hwan Bae, Jong Sung Park, Il-Young Oh,
Division of Cardiology, Department of Internal Medicine, Keimyung University Dongsan Hospital, Keimyung University School of Medicine, Daegu, Republic of Korea. · Division of Cardiology, Department of Internal Medicine, Severance Hospital, Yonsei University College of Medicine, Seoul, Republic of Korea.
Key points
From the paper's conclusion- In this exploratory single-arm study, appropriate ICD therapy was infrequent and exceeded by inappropriate shocks, while non-arrhythmic competing mortality dominated the clinical course. These hypothesis-generating findings support cautious patient selection, verification of filter status during follow-up, and larger comparative studies.
Abstract
BACKGROUND: Prospective data on the subcutaneous implantable cardioverter-defibrillator (S-ICD) in chronic kidney disease (CKD) are lacking. We estimated the rate of appropriate ICD therapy with the S-ICD in CKD stage 3-5 with left ventricular ejection fraction (LVEF) ≤ 35%. METHODS: In this single-arm exploratory study, adults with CKD stage 3-5 and a guideline-based primary-prevention indication received an EMBLEM MRI S-ICD at eight Korean centers. The primary endpoint, appropriate ICD therapy, was hypothesized at 18% versus a historical rate of 7.8% (α = 0.05, 80% power, planned N = 77; exact binomial test). RESULTS: The study was terminated early after 39 patients (51% of target; achieved power ≈61%). Mean LVEF was 27.8% and median eGFR 11.7 mL/min/1.73 m2; 71.8% had CKD stage 5 and 69.2% were on hemodialysis. Over a median follow-up of 361 days, appropriate ICD therapy occurred in 2 patients (5.1%; 95% CI, 1.4-16.9), significantly below the hypothesized 18% (p = 0.036). Inappropriate shocks occurred in 5 patients (12.8%; 95% CI, 5.6-26.7; 8 shocks), a proportion 2.5 times that of appropriate therapy, from external electromagnetic interference (2 patients), multiple counting after automatic SMART Pass deactivation (2), and supraventricular misclassification (1). Three deaths (7.7%) occurred, none device-related or documented as arrhythmic (one undetermined); all hospitalizations were for heart failure, and no systemic device infection occurred. CONCLUSIONS: In this exploratory single-arm study, appropriate ICD therapy was infrequent and exceeded by inappropriate shocks, while non-arrhythmic competing mortality dominated the clinical course. These hypothesis-generating findings support cautious patient selection, verification of filter status during follow-up, and larger comparative studies. TRIAL REGISTRATION: Clinical Research Information Service (CRIS), Republic of Korea; identifier KCT0005596.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Safety and efficacy of subcutaneous implantable cardioverter-defibrillator in dialysis patients: insights from a systematic reviewEuropean Heart Journal · 2026
- Real-world outcomes of subcutaneous ICD implants: a single-centre experienceEuropean Heart Journal · 2026
- Subcutaneous versus transvenous implantable cardioverter-defibrillator therapy: an updated systematic review and meta-analysis.Open heart · 2026
- Outcomes of Subcutaneous Implantable Cardioverter Defibrillators in Patients Being Considered for Heart Transplantation: A Retrospective Cohort Study.Pacing and clinical electrophysiology : PACE · 2026
- Long-term outcomes of subcutaneous implantable cardioverter defibrillators: the French nationwide HONEST studyEuropean Heart Journal · 2025
- Long term shock efficacy of the subcutaneous defibrillator in the EFFORTLESS extension sub studyEuropean Heart Journal · 2026