European Heart JournalMedicine

Long-term outcomes of subcutaneous implantable cardioverter defibrillators: the French nationwide HONEST study

Fawzi Kerkouri, Morin Christelle, Serge Bovéda, Frédéric Anselme, Soraya Anys, Nicolas Badenco, Francis Bessière, Hugues Blangy,

Inserm · Lille’s Cardiology Hospital

Observational · n=4,924

Key points

From the paper's conclusion
  • Overall, 4924 patients were enrolled (mean age 49.9 ± 15 years, 76.7% male, 63.0% for primary prevention).

Abstract

BACKGROUND AND AIMS: Although use of the subcutaneous implantable cardioverter defibrillator (S-ICD) is increasing, evidence from industry-independent and unselected populations remains limited. METHODS: HONEST is a ongoing nationwide academic observational study enrolling 98.2% of patients implanted with an S-ICD across France (2012-2019). Five-year clinical endpoints were centrally adjudicated. RESULTS: Overall, 4924 patients were enrolled (mean age 49.9 ± 15 years, 76.7% male, 63.0% for primary prevention). Implants used general anaesthesia (78.9%), and defibrillation testing (82.6%). Perioperative complications (within 30 days) occurred in 4.4%. At 5 years, cumulative incidence rates were 13.8% for inappropriate shocks, 10.8% for early battery depletion, 2.4% for infections, 1.5% for lead dysfunction, and 1.4% for chronic discomfort. Reoperation was required in 16.9%, need for cardiac pacing in 3.1%, and definite S-ICD extraction in 8.4%. Inappropriate shocks were independently associated with male sex (hazard ratio [HR] 1.29, 95% confidence interval [CI] 1.14-1.46, P < .001), obesity (HR 1.35, 95% CI 1.02-1.79, P = .032), arrhythmogenic right ventricular cardiomyopathy (HR 1.70, 95% CI 1.03-2.81, P = .036), and the presence of a pacemaker (HR 2.20, 95% CI 1.16-4.17, P = .016). SMART Pass filtering significantly reduced inappropriate shocks (HR 0.67, 95% CI 0.50-0.89, P = .007). Among patients with inappropriate shocks, ∼1% developed induced ventricular fibrillation (one fatality), and 10% underwent device extraction. Ineffective shocks or undetected arrhythmias occurred in only 0.2%. Among 547 deaths (11.1%), 53.9% were cardiovascular, including 26 sudden deaths, and 8 were S-ICD/procedure-related, with none related to S-ICD extraction. CONCLUSIONS: This nationwide study refines the long-term event profile of S-ICD therapy and may inform clinical practice and device selection.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Long-term outcomes of subcutaneous implantable cardioverter defibrillators: the French nationwide HONEST study — Science Skim