Validating a Wearable Bluetooth Oximeter for Infants in the Neonatal Intensive Care Unit.
Amy Courtney, Ju Lee Oei
Faculty of Medicine and Health, University of New South Wales, Sydney, New South Wales, Australia.
Key points
From the paper's conclusion- Wireless oximeters have poor correlation with wired oximeters especially at lower SpO2 levels but ease of use, especially in the home environment in less acutely ill patients, requires further evaluation.
Abstract
INTRODUCTION: Oxygen is vital for life but the wrong amount of oxygen can lead to significant risk of morbidity and mortality. Preterm infants are at particular risk from both too little (hypoxia) and excess (hyperoxia) oxygen. In the neonatal intensive care unit (NICU), oxygen treatment is guided by oximetry derived from wired oximeters which are difficult to use outside of the hospital environment. We aimed to validate Bluetooth wireless oximeters in newborn infants to determine the accuracy of devices that would be more suitable for out of hospital use. METHODS: Three 20 min sets of measurements including oxygen saturations (SpO2), heart rate (HR) and perfusion index (PI) from the post-ductal regions (left arm and lower limbs) were conducted in 21 spontaneously ventilating infants using the Masimo (1) SafetyNet and (2) Radical 7 oximeters. The wired Radical 7 oximeter was the reference. Data were downloaded onto CSV files and analysed for agreement. RESULTS: Twenty-one (eight male) infants were recruited between July to September 2023. Median gestations and ages were 36 weeks and 3 days. Neither Bluetooth device (SafetyNet and wireless Radical 7) met International Organization for Standardization (ISO) standard (ARMSE ≤ 3%), Score 3.624 (wireless Rad-7) and 3.884 (SafetyNet). Both wearables consistently overestimated SpO2, with mean biases (wired-wearable) of -1.802 (Rad-7) and -1.803 (SafetyNet), that increased at lower SpO2 levels. Linear regression, ICC, Pearson's correlation analyses showed poor consistency. CONCLUSION: Wireless oximeters have poor correlation with wired oximeters especially at lower SpO2 levels but ease of use, especially in the home environment in less acutely ill patients, requires further evaluation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparison of delivery room wrist versus earlobe pulse oximetry in newborns: A randomized controlled trial.Resuscitation · 2026
- Advanced Wireless Sensors for Neonatal Care in the Delivery Room - the AWARD Prospective International Study: A Research Protocol.JMIR research protocols · 2026
- AI-Driven and Automated Systems for Continuous Oxygen Saturation Monitoring in Long-Term Oxygen Therapy: Systematic Review.JMIRx med · 2026
- Development and validation of an early prediction model for post-discharge home oxygen therapy requirement in preterm infants.Respiratory medicine · 2026
- Design and pilot clinical evaluation of a low-cost, continuous respiratory monitor for newbornsmedRxiv · 2026
- P017. Agreement between two commonly used pulse oximeters during in-laboratory polysomnographySLEEP advances · 2026