Impact of Real-Time Tissue Proximity Indicator Visualization and Pulsed Field Tag Coloring on Lesion Delivery and Procedural Workflow During Atrial Fibrillation Ablation Using the VARIPULSE Catheter.
Nadica Trajkovska, Moneeb Khalaph, Angeliki Darma, Mustapha El Hamriti, Thomas Fink, Vanessa Sciacca, Denise Guckel, Maximillian Mörsdorf,
Department of Electrophysiology, Heart and Diabetes Center NRW, Ruhr University Bochum, Bochum, Germany. · Clinic for Cardiology/Electrophysiology, GZO-Spital, Wetzikon, Switzerland.
Key points
From the paper's conclusion- These findings suggest modified, proximity-guided lesion delivery without a relevant workflow penalty.
Abstract
INTRODUCTION: The updated VARIPULSE™ software (V8.1.1.944) introduces a real-time tissue proximity indicator, enhanced VISITAG™ differentiation of complete versus incomplete applications, and grid-based color coding for lesion assessment during pulsed field ablation (PFA). Its impact on procedural workflow and lesion delivery remains insufficiently defined. METHODS AND RESULTS: We compared 55 pulmonary vein isolation procedures performed with the updated software and 91 procedures performed with the previous version (V8.1.0.325) in 146 consecutive patients with atrial fibrillation. The updated software was associated with a higher number of PFA applications per procedure (19.76 ± 4.10 vs. 17.74 ± 3.90; p < 0.001), with more applications across all pulmonary veins, without increases in fluoroscopy time, radiation dose, or procedure duration (54.06 ± 14.99 vs. 56.64 ± 17.23 min; p = 0.046). Procedure time decreased further after an initial learning curve of 30 cases. First-pass isolation of all pulmonary veins was achieved in 83.6% versus 78.0% of procedures (p = 0.41). CONCLUSION: These findings suggest modified, proximity-guided lesion delivery without a relevant workflow penalty.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Procedural characteristics and acute outcomes with different pulsed field ablation systems for atrial fibrillationJournal of Interventional Cardiac Electrophysiology · 2026
- Feasibility of tissue proximity indication during atrial fibrillation rhythm as an indicator of acute pulmonary vein reconnection using variable-loop pulsed field ablation catheter.International journal of cardiology · 2026
- Non-complete tissue proximity indication shows perfect concordance with acute isolation gaps: an eight-segment spatial analysis during pulsed field ablation with a variable-loop circular catheterEuropean Heart Journal · 2026
- Clinical validation of a ≥35% contact index threshold for fixed-loop circular pulsed field ablationEuropean Heart Journal · 2026
- Development and Validation of a Novel Contact Force-Integrated Pulsed Field Quantization Index for Pulmonary Vein Isolation in a Porcine Model.Journal of cardiovascular electrophysiology · 2026
- PVI Using the Fixed Loop PFA in Paroxysmal AF: One-Year Outcomes with Continuous Implantable Loop Recorder Monitoring.Indian pacing and electrophysiology journal · 2026