Atrial Pacing Capability of a Single-Lead Implantable Cardioverter-Defibrillator Incorporating an Atrial Floating Dipole.

Manuel Cerini, Gianmarco Arabia, Gioele Papi, Paolo Fornaro, Emiliano Calvi, Alessio Nicolai, Aurora Giavarini, Gianfranco Mitacchione,

Cardiology Department, Ospedale di Cremona, Cremona, Italy. · Department of Electrophysiology and Cardiac Pacing, ASST-Spedali Civili, Department of Medical and Surgical Specialties, Radiological Sciences and Public Health, University of Brescia, Brescia, Italy.

Observational

Key points

From the paper's conclusion
  • Atrial capture through the floating atrial dipole of a single-lead DX ICD system is feasible in the majority of patients, with bipolar pacing providing the highest capture rate and the lowest PTs. These findings provide proof-of-concept evidence supporting further development of single-lead dual-chamber sensing and pacing strategies using the DX platform.

Abstract

INTRODUCTION: The single-lead DX implantable cardioverter-defibrillator (ICD) combines the advantages of a conventional single-lead ICD with atrial sensing capability provided by an integrated atrial floating dipole. While atrial sensing performance has been extensively documented, the ability of the floating atrial dipole to achieve atrial capture has not previously been investigated. METHODS AND RESULTS: Consecutive patients undergoing de novo implantation or generator replacement of a single-lead DX ICD system at two Italian centers were prospectively enrolled. Acute atrial pacing was assessed using three configurations: bipolar pacing and two unipolar configurations employing either the proximal or distal atrial electrode. Atrial capture, pacing threshold (PT), impedance, and phrenic nerve stimulation were evaluated at pulse widths of 0.4 and 1.0 ms. Twenty-five patients (92% male; mean age 66.2 ± 10.1 years) were included. Acute atrial capture was achieved in 22 patients (88%) with bipolar pacing, compared with 18 (72%) and 17 (68%) using proximal and distal unipolar pacing, respectively (p = 0.122). Bipolar pacing yielded significantly lower PTs than unipolar configurations at both pulse widths (p < 0.020). Median atrial sensing amplitude measured after connection to the ICD generator was 3.9 mV (2.8-5.1). CONCLUSION: Atrial capture through the floating atrial dipole of a single-lead DX ICD system is feasible in the majority of patients, with bipolar pacing providing the highest capture rate and the lowest PTs. These findings provide proof-of-concept evidence supporting further development of single-lead dual-chamber sensing and pacing strategies using the DX platform.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Atrial Pacing Capability of a Single-Lead Implantable Cardioverter-Defibrillator Incorporating an Atrial Floating Dipole — Science Skim