Consensus-Recommended Follow-Up Mediates Cardiovascular Outcomes Disparities Between Implantable Cardioverter Defibrillator and Pacemaker Recipients.
Jingwen Huang, Rahul Patel, Chang Liu, Michael S Lloyd
Section of Cardiac Electrophysiology, Department of Medicine, Emory University School of Medicine, Atlanta, Georgia, USA. · Department of Epidemiology, Emory University Rollins School of Public Health, Atlanta, Georgia, USA.
Key points
From the paper's conclusion- More than half of the excess cardiovascular risk in ICD recipients was statistically mediated by suboptimal follow-up adherence-a potentially modifiable pathway. These findings identify follow-up adherence as a potentially important pathway and support further prospective investigation of strategies to improve monitoring and clinical outcomes in high-risk CIED populations.
Abstract
BACKGROUND: Consensus-recommended follow-up after cardiac implantable electronic device (CIED) implantation improves outcomes, yet whether it mediates cardiovascular disparities between device types remains unknown. METHODS: We prospectively studied 6806 patients undergoing CIED implantation across a multi-hospital system (2019-2023; median follow-up 3.5 years), categorized as implantable cardioverter defibrillator (ICD) (n = 2757) or permanent pacemaker (PPM) (n = 4049). Causal mediation analysis decomposed the effect of device type on major adverse cardiovascular events (MACE) and device extraction into direct and follow-up mediated components, adjusting for demographics and comorbidities. RESULTS: ICD recipients had significantly lower consensus-recommended follow-up rates than PPM recipients (65.3% vs. 79.2%; adjusted OR 0.46, 95% CI 0.40-0.52, p < 0.001) and substantially higher MACE risk (HR 1.89, 95% CI 1.66-2.25, p < 0.001), of which 53.5% was mediated by follow-up adherence. Device extraction rates were lower in ICD recipients (HR 0.30, 95% CI 0.24-0.38), with only 7.9% mediated by follow-up adherence. CONCLUSIONS: More than half of the excess cardiovascular risk in ICD recipients was statistically mediated by suboptimal follow-up adherence-a potentially modifiable pathway. These findings identify follow-up adherence as a potentially important pathway and support further prospective investigation of strategies to improve monitoring and clinical outcomes in high-risk CIED populations.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Long-term incidence of infective endocarditis according to the type of cardiac implantable electronic device: a nationwide comparison with moderate- and high-risk patient groups.Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology · 2026
- Primary Prevention ICD in a Contemporary Underserved Population: Factors Associated With Underutilization and Mortality.Pacing and clinical electrophysiology : PACE · 2026
- Predicting Major Adverse Events With Cardiac Resynchronization, Leadless, and Conduction System Pacemaker Implantation.Pacing and clinical electrophysiology : PACE · 2026
- Barriers to cardiac device access in heart failure patients: a multicenter study in Mexico.Archivos de cardiologia de Mexico · 2026
- Cardiac Implantable Electronic Devices Management in Older Adults ≥75 Years Old: From Underrepresentation in Clinical Trials to Overrepresentation in Contemporary Clinical Practice.Current cardiology reviews · 2026
- Predictors of Appropriate Implantable Cardioverter-Defibrillator Intervention in Adults with Congenital Heart Disease.Journal of clinical medicine · 2026