Evaluation of Reticulocyte Channel on Sysmex XN ‐10 for Resolving Cold Agglutination in Samples With High Prevalence of Thalassemic Hemoglobinopathies
Shir Ying Lee, Jia Hui Goh, Ming Zi Soh, Jin De Leong, Florbie S. Pedro, Stephrene S. W. Chan
Key points
From the paper's conclusion- This study evaluated the Sysmex XN‐10 reticulocyte channel for resolving cold agglutination and assessed the performance of delta RBC and delta MCHC in a population with a high prevalence of thalassemia and hemoglobinopathies (TH).
Abstract
Introduction Cold agglutinins interfere with automated complete blood count analysis by causing red cell aggregation and spurious erythrocyte indices. This study evaluated the Sysmex XN‐10 reticulocyte channel for resolving cold agglutination and assessed the performance of delta RBC and delta MCHC in a population with a high prevalence of thalassemia and hemoglobinopathies (TH). Methods Twenty healthy controls, 29 TH samples, and 126 consecutive specimens with MCHC > 38.5 g/dL were studied. Optical parameters obtained in the preincubation reticulocyte channel were compared with conventional impedance parameters after incubation at 37°C. Agreement, correlation, and receiver operating characteristic analyses were performed. Results Optical and impedance parameters showed strong agreement in healthy and TH cohorts. Of 126 high‐MCHC specimens, 122 showed cold agglutination on smear; 84 had quantifiable preincubation values, and 72 normalized after incubation and were included for quantitative comparison. Preincubation optical parameters showed strong concordance with postincubation impedance values. Twelve samples with persistent postwarming abnormalities appeared better resolved in the reticulocyte channel. Delta RBC % ((RBC‐O − RBC‐I)/RBC‐I × 100) was highly sensitive for cold agglutination, with 100% sensitivity at a threshold of > 5%, while specificity decreased in TH‐prevalent cohorts. Delta MCHC % ((cMCHC − MCHC)/MCHC × 100) showed excellent discriminatory performance, with an optimal cut‐off of < −7.7% yielding 98.8% sensitivity and 100% specificity. Conclusion The Sysmex XN‐10 reticulocyte channel provides a reliable approach for resolving cold agglutination‐related interference in complete blood count analysis. In TH‐prevalent settings, sequential use of delta RBC and delta MCHC potentially improves discrimination and facilitates more efficient release of reliable hematology results.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Rapid Electrochemical Impedance Detection of the Mia Antigen on Red Blood Cells.ACS omega · 2026
- Methemoglobin Causes Collapsed WDF Scattergrams and a Low-SFL Mature RBC Population in RET Scattergrams on the XR-Series Automated Hematology Analyzer.International journal of laboratory hematology · 2026
- A-229 Usefulness of Reticulocyte Hemoglobin Expression (RHE) Integrated into the Blood Count in the Differential Diagnosis of Iron Deficiency Anemia and Anemia of InflammationClinical Chemistry · 2026
- The Presence and Percentage of Circulating Nucleated Red Blood Cells Reveal Distinct Characteristics in Adults With Sickle Cell Disease.American journal of hematology · 2026
- Deoxygenation-free digital holographic microscopy discriminates sickle-cell disease in a piloted Colombian cohort via phase and shape metricsOptics Express · 2025
- CD71-based immature RBC (iRBC) analysis in peripheral blood has a better correlation with leukocyte PNH clone size and improves the delineation of different PNH clone types.Cytometry. Part B, Clinical cytometry · 2026