Bench-Top Evaluation of Pancreatobiliary Guidewires for Endoscopic Ultrasound-Guided Drainage.

Takehiko Koga, Yusuke Ishida, Naoaki Tsuchiya, Takanori Kitaguchi, Keisuke Matsumoto, Makoto Fukuyama, Tatsuro Nii, Norihiro Kojima,

Department of Gastroenterology and Medicine, Fukuoka University Faculty of Medicine, Fukuoka, Japan. · Department of Gastroenterology, Taragi Municipal Hospital, Kumamoto, Japan.

Key points

From the paper's conclusion
  • This bench-top study suggests that structural parameters related to the shaft and tip are associated with guidewire behavior during EUS-D/A.

Abstract

OBJECTIVES: Guidewire manipulation is a key determinant of procedural success in endoscopic ultrasound-guided drainage/anastomosis (EUS-D/A). However, most guidewires were originally designed for endoscopic retrograde cholangiopancreatography, and objective selection criteria for EUS-D/A have not been established. This study aimed to clarify the relationships between the structural features and the mechanical performance of guidewires used in EUS-D/A. METHODS: Twenty-one commercially available guidewires (0.018-0.035 in.) were evaluated. Structural parameters, including the polyurethane length, tip deflection height, tip and shaft core wire diameters, outer diameter, covering texture, and covering adhesion, were measured. Mechanical performance was assessed using bench-top experiments including insertion and withdrawal forces, torque response and whip motion within an endoscopic ultrasound-guided tissue acquisition (EUS-TA) needle, bending force, tip load, loop performance, and the turning point of loop formation. Correlations between the structural and performance parameters were analyzed. RESULTS: Core wire diameter exhibited a strong positive correlation with the bending force, which was strongly correlated with the insertion and withdrawal forces. Tip core diameter was strongly correlated with the tip load, and loop formation consistently occurred at the initiation of the steeply tapered core wire segment. Torque response rate was positively correlated with the core wire diameter and core-to-outer diameter ratio. Guidewires with a core wire diameter ≥ 0.52 mm and polytetrafluoroethylene-based gentle-pitch uneven coverings demonstrated a higher torque response rate. CONCLUSIONS: This bench-top study suggests that structural parameters related to the shaft and tip are associated with guidewire behavior during EUS-D/A.

Read the paper