Journal of the mechanical behavior of biomedical materials

Quantification of the flexural rigidity of endovascular surgical devices using three-point bending tests.

Michael Y Qiu, Charles B Suskin, Juan J Becerra-Garcia, Sophia R Pyeatte, DeVaughn G Rucker, Mohamed A Zayed, Joshua W Osbun, Guy M Genin

CardioVascular Research Innovation in Surgery & Engineering Center, Washington University in St. Louis, St. Louis, MO, USA; Division of Neurotechnology, Department of Neurological Surgery, Washington University School of Medicine, St. Louis, MO, USA; Department of Biomedical Engineering, Washington University in St. Louis, St. Louis, MO, USA. · CardioVascular Research Innovation in Surgery & Engineering Center, Washington University in St. Louis, St. Louis, MO, USA; Division of Neurotechnology, Department of Neurological Surgery, Washington University School of Medicine, St. Louis, MO, USA; Department of Mechanical Engineering and Material Sciences, Washington University in St. Louis, St. Louis, MO, USA; Center for Innovation in Neuroscience and Technology, Washington University School of Medicine, St. Louis, MO, USA; Division of Vascular Surgery, Department of Surgery, Washington University School of Medicine, St. Louis, MO, USA; NSF Science and Technology Center for Engineering Mechanobiology, Washington University in St. Louis, St. Louis, MO, USA.

Key points

From the paper's conclusion
  • We therefore developed a three-point bend test methodology using a custom rig and applied it to measure lengthwise flexural rigidity profiles of devices commonly used in neuroendovascular procedures.

Abstract

Endovascular surgical procedures require the navigation of catheters and wires through the vasculature to reach distal target sites. Quantitative frameworks for device selection have the potential to improve navigation success by optimizing device flexural rigidity for a given patient-specific anatomy. However, data are lacking to facilitate this technology, in part because typical endovascular devices have intricate spatial variations in mechanical properties that are challenging and tedious to quantify repeatably. We therefore developed a three-point bend test methodology using a custom rig and applied it to measure lengthwise flexural rigidity profiles of devices commonly used in neuroendovascular procedures. The methodology demonstrated high repeatability and was able to characterize transition zones in detail. We applied the methodology to generate the first comprehensive, quantitative library of device flexural rigidities, spanning guidewires, intermediate guides, and long sheaths. We observed that these three classes of device have properties that fall into distinct ranges. Additional plots examining relationships between flexural rigidity, device diameter, and length revealed application-specific trends in flexural properties. This methodology and resulting data allow for standardized characterization and comparisons to aid device selection, and have the potential to both enhance surgical planning and inform future endovascular innovation.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers