First Cases of Allergic Contact Dermatitis From the New FreeStyle Libre 3 Sensor: Small Device, Multiple Allergens.
Silada Kanokrungsee, Ella Dendooven, Kenn Foubert, Tania Naessens, Emmy Tuenter, Nina Hermans, Olivier Aerts
Department of Dermatology, University Hospital Antwerp (UZA) and Research Group Immunology, INFLA-MED Centre of Excellence, University of Antwerp, Antwerp, Belgium. · Natural Products and Food Research and Analysis - Pharmaceutical Technology, Department of Pharmaceutical Sciences, University of Antwerp, Antwerp, Belgium.
Key points
From the paper's conclusion- Multiple newly identified as well as (modified) classic sensitizers contribute to ACD from the FreeStyle Libre 3 glucose sensor. Comprehensive ingredient disclosure and extended patch testing are essential for accurate diagnosis and prevention.
Abstract
BACKGROUND: Allergic contact dermatitis (ACD) from glucose sensors has become increasingly common. Newer devices, such as FreeStyle Libre 3 (Abbott Diabetes Care, UK), have been developed to enhance performance and reduce skin reactions. No ACD cases linked to this particular device have been reported to date. OBJECTIVE: To document the first cases of ACD linked to the FreeStyle Libre 3 sensor and to identify potential culprit allergenic components through patch testing and chemical analyses. METHODS: A retrospective chart review was conducted on 15 patients with suspected ACD related to the FreeStyle Libre 3 device who underwent patch testing at Antwerp University Hospital between November 2023 and November 2025. Chemical analyses of the device and associated adhesive components were performed using gas chromatography-mass spectrometry (GC-MS). RESULTS: Ten patients (67%) were diagnosed with ACD and most had prior exposure to other glucose sensors. The most frequent allergens were colophonium and related compounds, followed by (meth)acrylates, isocyanates and tert-butylphenol (TBP) derivatives. GC-MS analyses revealed the presence of most of these chemicals. CONCLUSIONS: Multiple newly identified as well as (modified) classic sensitizers contribute to ACD from the FreeStyle Libre 3 glucose sensor. Comprehensive ingredient disclosure and extended patch testing are essential for accurate diagnosis and prevention.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- High Diagnostic Yield of Testing With a Diabetes Device Screening Series in Patients With Suspected Allergic Reactions to Glucose Sensors, Insulin Patch Pumps and Insulin Infusion Sets.Contact dermatitis · 2026
- Prevalence of Contact Sensitization in Patients with Atopic Dermatitis: A Retrospective Study from China.Acta dermato-venereologica · 2026
- Sensitization potential of chemical constituents in earphone materials in high-risk populations: a multicenter patch test study.PeerJ · 2026
- North American Contact Dermatitis Group Patch Test Results: 2023-2024.Dermatitis : contact, atopic, occupational, drug · 2026
- Assessing the Skin Sensitizing Potential of Constituents and Leachables from Wearable Devices Using GARDskin and Associated Applications.Toxics · 2026
- Verification of a Quantitative Risk Assessment Approach for Skin Sensitization Using EC1.6 Derived From LLNA: BrdU-ELISA With Medical Devices.Journal of applied toxicology : JAT · 2026