Compound Lens Design and Beamforming for Wide-Field Volumetric Imaging Using a Row-Column Array.
Ali Salari, Melanie Audoin, Andre Rath, Erik Vilain Thomsen, Jorgen Arendt Jensen
Key points
From the paper's conclusion- These results demonstrate that compound-lens RCA systems, combined with accurate and efficient beamforming, enable wide-field volumetric imaging of organ-scale structures while preserving the practical advantages of RCA arrays, with a theoretical volume rate of approximately 104 Hz, indicating the potential for real-time 3-D visualization.
Abstract
Row-column addressed (RCA) ultrasound arrays enable volumetric imaging with a significantly reduced channel count compared to fully addressed 2-D arrays; however, their field of view (FOV) is typically limited to the transducer footprint. Diverging acoustic lenses can extend the FOV, but accurate beamforming remains challenging when the lens's finite thickness and multilayer structure are not properly modeled. This work presents a compound acoustic lens design that simultaneously achieves a large FOV, reduced edge-wave artifacts, and acceptable sensitivity loss, together with a beamforming approach that accurately accounts for wave propagation through multilayer lens structures. The proposed approach combines ray tracing with a Normal Path (RT-NP) approximation and determines the optimal propagation path using a two-dimensional Newton-Raphson optimization, enabling accurate time-of-flight (TOF) estimation while significantly reducing computational complexity compared to conventional two-interface ray-tracing methods. The method is validated experimentally using a 128+128 Vermon RCA transducer operating at 6 MHz and equipped with a BiQuadratic compound lens composed of TPX and RTV-PTFE materials. Pressure-field measurements validate the predicted wavefront expansion, while simulations indicate an FOV expansion to approximately 54.5°, enabling imaging well beyond the transducer aperture while preserving high spatial resolution and image quality. Hydrophone and RF-data measurements demonstrate sub-wavelength TOF accuracy of the proposed beamformer. The proposed system achieves a penetration depth of 75 mm compared with 90 mm without the lens, with only minor degradation in spatial resolution. These results demonstrate that compound-lens RCA systems, combined with accurate and efficient beamforming, enable wide-field volumetric imaging of organ-scale structures while preserving the practical advantages of RCA arrays, with a theoretical volume rate of approximately 104 Hz, indicating the potential for real-time 3-D visualization.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Angular-first channel-domain delay-multiply-and-sum beamforming with dual coherence for row-column array-based 3D power Doppler imaging.Ultrasonics · 2026
- Beamforming Through Encapsulation Layers: Efficient Time-of-Flight Modeling for High-Resolution Ultrasound Imaging.IEEE transactions on ultrasonics · 2026
- Bias-Switchable Row-Column Array Imaging Using Fast Orthogonal Row-Column Electronic Scanning (FORCES) Compared With Conventional Row-Column Array Imaging.IEEE open journal of ultrasonics, ferroelectrics, and frequency control · 2026
- A 240-Element Matrix Probe With Aberration Mask for 4-D Carotid Artery Computational Ultrasound Imaging.IEEE transactions on ultrasonics · 2026
- Generalized Apodization Design with Learned Complex-Valued Weights for High-Frame-Rate Ultrasound Beamforming.IEEE transactions on ultrasonics · 2026
- Four-Dimensional Sequential Elastography for Ocular Tissue Using 2D Matrix Array.BME frontiers · 2026