A Parameterization Framework for Image-Based Resolution Modeling of Spatially-Variant PSF Deformation in Dedicated Breast PET.
Min Gao, Srilalan Krishnamoorthy, Juhi Raj, Samuel Matej, Suleman Surti
Department of Radiology, Perelman School of Medicine, University of Pennsylvania, Philadelphia, PA 19104 USA.
Key points
From the paper's conclusion- This work provides a practical strategy for incorporating spatially-variant resolution modeling into limited-angle PET systems and demonstrates its potential to improve quantitative imaging performance in dedicated breast PET applications.
Abstract
Spatially-variant (SV) point spread function (PSF) deformation in limited-angle PET systems can significantly degrade lesion localization, shape fidelity, and quantitative accuracy, particularly in dual-panel breast PET (BPET) scanners. While image-based resolution modeling (IRM) has been shown to improve image quality, scanner implementation of SV-IRM remains a major challenge, as it is not practical to measure hundreds of PSFs, as used in previous simulation studies. In this work, we propose a parameterization framework to characterize SV PSF deformation using a limited number of measured PSFs. Each PSF kernel is modeled using a Gaussian mixture model (GMM), and the spatial variation of the GMM parameters is described using polynomial fitting across the field of view. This approach enables the generation of a dense PSF lookup table from a modest number of measurements, allowing practical implementation of SV-IRM in a real BPET system. The proposed method was evaluated using multiple physical phantoms, including hot-rod, single-lesion, and NEMA NU 4 phantoms. Results demonstrate that SV-IRM improves lesion localization, restores elongated structures, and achieves higher contrast recovery compared to both no IRM and spatially-invariant IRM approaches. A moderate increase in background noise was observed, reflecting the inherent trade-off between resolution recovery and noise amplification. This work provides a practical strategy for incorporating spatially-variant resolution modeling into limited-angle PET systems and demonstrates its potential to improve quantitative imaging performance in dedicated breast PET applications.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Impact of depth-of-interaction and spatially-variant point-spread-function model on the performance of ultra-high resolution brain PET.Physics in medicine and biology · 2026
- Deep unrolled regularization for limited-angle dual-panel positron emission mammography.Physical and engineering sciences in medicine · 2026
- Design and imaging performance of BPET-DBT - a dedicated scanner for breast imaging.Medical physics · 2026
- Multi-GPU parallel framework for DOI-enabled long axial field-of-view PET image reconstruction.Physics in medicine and biology · 2026
- Lesion-Aware Adaptive Fourier Neural Operator for CT-to-PSMA PET Synthesis in Prostate CancerarXiv · 2026
- Sinogram-based regularization parameter learning for 2D penalized maximum-likelihood PET image reconstruction.Physics in medicine and biology · 2026