Physics in medicine and biology

Impact of depth-of-interaction and spatially-variant point-spread-function model on the performance of ultra-high resolution brain PET.

Tianyi Zeng, Yassir Najmaoui, Jiazhen Zhang, Tommaso Volpi, Kathryn Fontaine, Yanis Chemli, Yifan Hu, Jean-Dominique Gallezot,

Department of Radiology and Biomedical Imaging, Yale School of Medicine, 330 Cedar St, New Haven, Connecticut, 06520, United States. · Yale School of Medicine Department of Radiology & Biomedical Imaging, 330 Cedar St, New Haven, Connecticut, 06520, United States.

Key points

From the paper's conclusion
  • The findings motivate combining modest DOI capability with accurate resolution modeling, while further validation is required for redesigned detectors and fully corrected acquisitions.

Abstract

To implement a spatially-variant point spread function (VarPSF) model for the NeuroEXPLORER (NX) brain positron emission tomography (PET) scanner in the open-source Yale Reconstruction Toolkit for PET (YRT-PET), and investigate the interplay between depth-of-interaction (DOI) resolution and resolution modeling. Approach. PSF parameters were estimated from simulated and physically measured point sources. Three primary DOI configurations were evaluated: eight-layer DOI (8DOI), two-layer DOI (2DOI), and no DOI (noDOI), with an additional 4DOI comparison in the off-center simulated Derenzo study. Reconstructions using VarPSF, a spatially-invariant PSF measured at the field-of-view (FOV) center (SinglePSF), and no PSF modeling (noPSF) were assessed using simulated Derenzo and brain phantoms, a physical Derenzo phantom, and one non-human primate (NHP) scan. Main results. In simulated and physical Derenzo studies, VarPSF yielded higher peak-to-valley ratios (PVRs) than SinglePSF and noPSF at the final evaluated iterations, with pronounced gains at large radial offsets. In the simulated brain phantom, VarPSF reduced the estimated transverse blur by 0.3 mm relative to SinglePSF and 0.5 mm relative to noPSF in 2DOI, and by 0.6 mm and 0.8 mm, respectively, in noDOI. In the NHP study, VarPSF improved contrast and structural delineation with reduced DOI encoding at approximately matched white-matter coefficient of variation, while its incremental benefit over SinglePSF was small with 8DOI at clinically relevant iteration settings. Significance. Spatially-variant PSF modeling can partially compensate for limited DOI information. The findings motivate combining modest DOI capability with accurate resolution modeling, while further validation is required for redesigned detectors and fully corrected acquisitions. The open-source implementation enables controlled investigation of the complementary roles of DOI encoding and resolution modeling.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers