Design and imaging performance of BPET-DBT - a dedicated scanner for breast imaging.
Srilalan Krishnamoorthy, Emmanuel Morales, William J Ashmanskas, Matthew E Werner, Samuel Matej, Juhi Raj, Trevor L Vent, Chloe J Choi,
Department of Radiology, University of Pennsylvania, Philadelphia, Pennsylvania, USA. · Department of Physics and Astronomy, University of Pennsylvania, Philadelphia, Pennsylvania, USA.
Key points
From the paper's conclusion- We successfully constructed and tested a BPET scanner and integrated it with DBT. BPET scanner performance was characterized, and imaging capabilities relevant for clinical imaging have been highlighted. Images from a patient with ER+ breast cancer show clear tumor delineation and visualization of heterogeneous uptake, as also seen with WB-PET. This approach offers a promising strategy for improved treatment planning and outcomes in breast cancer care.
Abstract
BACKGROUND: A dedicated breast PET (BPET) scanner integrated with digital breast tomosynthesis (DBT) enables three-dimensional functional imaging with quantitative information co-registered to high-resolution anatomical images. With the availability of breast cancer specific PET radiotracers, such a scanner has the potential to improve treatment planning in breast cancer. We recently developed a BPET-DBT scanner utilizing an integrated gantry design to generate intrinsically co-registered BPET-DBT images. PURPOSE: This paper presents the scanner design, characterizes its imaging performance, and reports on our initial human imaging experience to demonstrate scanner's clinical readiness. METHODS: The PET component is comprised of two detector heads, providing a 20 × 10 × 10 cm3 scanner field-of-view. The detector provides time-of-flight (TOF) measurement and is comprised of 32×32 LYSO arrays of 1.5 × 1.5 × 15 mm3 crystals coupled to multi-anode PMTs. Custom waveform-sampling electronics ensures high spatial and timing resolution with minimal deadtime. The DBT component is a state-of-the-art system which employs 2D x-ray tube and detector motion for improved contrast and resolution. Phantoms appropriate for breast imaging were imaged to assess spatial resolution, image quantitation, count-rate capability, and co-registration accuracy. First human imaging was performed in volunteers with estrogen-receptor-positive (ER+) breast cancer. A five-minute BPET-DBT scan using 18F-fluoroestradiol was acquired and compared with a whole-body PET scan. RESULTS: System timing resolution of 425 ps and energy resolution of 16% were measured. Spatial resolution measurements showed an in-plane (parallel to the detector) resolution of 2 mm and demonstrates the ability to resolve 1.6 mm rods in the micro-deluxe hot rod phantom. TOF reconstruction combined with image-based resolution modeling mitigates artifacts typical in scanners with incomplete angular coverage and enables good out-of-plane (orthogonal to detector) discrimination of the 2.4 mm rods. Count-rate measurements confirm low scanner dead-time, with sufficient count-rate capability for clinical use. CONCLUSIONS: We successfully constructed and tested a BPET scanner and integrated it with DBT. BPET scanner performance was characterized, and imaging capabilities relevant for clinical imaging have been highlighted. Images from a patient with ER+ breast cancer show clear tumor delineation and visualization of heterogeneous uptake, as also seen with WB-PET. This approach offers a promising strategy for improved treatment planning and outcomes in breast cancer care.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Development and performance evaluation of an upright dedicated cone-beam breast CT system.Medical physics · 2026
- Impact of depth-of-interaction and spatially-variant point-spread-function model on the performance of ultra-high resolution brain PET.Physics in medicine and biology · 2026
- A Parameterization Framework for Image-Based Resolution Modeling of Spatially-Variant PSF Deformation in Dedicated Breast PET.IEEE transactions on radiation and plasma medical sciences · 2026
- Design and Development of NeuroSphere PET: A Next-Generation 7-T MR-Compatible Dedicated Brain PET Scanner.Journal of nuclear medicine : official publication, Society of Nuclear Medicine · 2026
- Development of a multifunctional anthropomorphic phantom for performance evaluation in nuclear medicine imaging: preliminary PET/CT validation.EJNMMI physics · 2026
- Performance of Photon-counting CT for Assessing Pretreatment Breast Cancer: Comparison with Mammography, MRI, and 18F-FDG PET/CT.Radiology · 2026