Acute Stimulation Selectivity of Minimally Invasive Interfascicular Nerve Stimulation (MIINS) in a Large Animal Model (June 2026).
Margaux A Randolph, Vlad I Marcu, Michael J Rinderle, Constantinos Tsiptsis, Andrew J Shoffstall, Jonathan Z Baskin, Dustin J Tyler
Key points
From the paper's conclusion- Acute muscle selectivity and micro-CT characterization of electrode placement in the median nerves of Yorkshire pigs demonstrated MIINS's feasibility and capability to achieve high selectivity and distributed interfascicular contact placement. MIINS achieved selective muscle recruitment comparable to that of the C-FINE and other existing intrafascicular designs in the literature. The data also show that in large nerves, higher anatomical invasiveness does not guarantee superior selectivity.
Abstract
OBJECTIVE: Interfascicular peripheral nerve interfaces may offer a balance of high selectivity and reduced invasiveness across peripheral nerve stimulation applications. This study compares the stimulation selectivity of the Minimally Invasive Interfascicular Nerve Stimulation (MIINS) interface with that of the extraneural Composite Flat Interface Nerve Electrode (C-FINE) in a porcine model. Microcomputed tomography (micro-CT) quantified the placement of MIINS contacts relative to the fascicles and to the other contacts within the nerve. METHODS: Muscle selectivity was assessed with electromyography (EMG) across 13 MIINS and 7 C-FINE trials in the median nerves of Yorkshire pigs. Micro-computed tomography characterized 126 total MIINS microwire placements across 26 nerves of the brachial plexus. RESULTS: Both devices achieved muscle selectivity: C-FINE selectively activated 3.00 ± 0.82 muscles, and MIINS selectively activated 2.38 ± 0.96 muscles out of the 5 total muscles sampled. MIINS achieved this with fewer contacts. Micro-CT confirmed that 89% of MIINS microwires were placed interfascicularly, with the majority (74%) localized to the central half of the nerve cross-section. SIGNIFICANCE: Acute muscle selectivity and micro-CT characterization of electrode placement in the median nerves of Yorkshire pigs demonstrated MIINS's feasibility and capability to achieve high selectivity and distributed interfascicular contact placement. MIINS achieved selective muscle recruitment comparable to that of the C-FINE and other existing intrafascicular designs in the literature. The data also show that in large nerves, higher anatomical invasiveness does not guarantee superior selectivity.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Selective activation of nerve fiber subpopulations with intrafascicular stimulation.Journal of neural engineering · 2026
- STRIDE: A Flexible, Depth-Controlled Intramuscular EMG Platform for Small-Animal Studies.IEEE transactions on neural systems and rehabilitation engineering : a publication of the IEEE Engineering in Medicine and Biology Society · 2026
- Using a single penetrating interfascicular electrode to improve spatial selectivity of an extraneural polymeric cuff arrayBioelectronic Medicine · 2025
- An open-source thin film extraneural cuff electrode for peripheral nerve stimulation.IEEE transactions on bio-medical engineering · 2026
- Multi-scale computational modeling of magnetoelectric nanoparticle-dispersed conductive nanofibers for peripheral nerve modulation.Frontiers in bioengineering and biotechnology · 2026
- Chronic Tibial Nerve Stimulation Augmented by Injection of an Easy-to-Remove Helical Wire Structure Electrode in Swine.Neuromodulation : journal of the International Neuromodulation Society · 2026