Multi-scale computational modeling of magnetoelectric nanoparticle-dispersed conductive nanofibers for peripheral nerve modulation.
Valentina Galletta, Serena Fiocchi, Claudio Verardo, Luigi Carbone, Loretta Laureana Del Mercato, Simone Romeni, Emma Chiaramello
Department of Electronics, Information and Bioengineering (DEIB), Politecnico di Milano, Milan, Italy. · Institute of Electronics, Computer and Telecommunication Engineering (IEIIT), National Research Council (CNR), Milan, Italy.
Key points
From the paper's conclusion- The solution of embedding MENPs in conductive polymer nanofibers enables higher MENPs loading than direct tissue injection, improves positional stability, and preserves biocompatibility, thereby offering a potential alternative for the engineering of MENPs as neuromodulation interfaces.
Abstract
INTRODUCTION: The electrical stimulation of peripheral nerves has shown the potential to restore impaired functions across a wide range of clinical scenarios. In most cases, though, the lack of selectivity of existing stimulation interfaces limits their applicability and can in some cases produce severe adverse effects. In addition, traditional stimulation requires physical connectors to a pulse generator that can add further biological and mechanical stress already linked to interfacing with physiological structures. This study focuses on the use of magnetoelectric nanoparticles (MENPs) embedded within conductive polymeric nanofibers as new wireless, highly selective, and minimally invasive interfaces for modulating peripheral nerves activity. METHODS: A step-by-step detailed characterization of the physical variables involved in this interaction, spanning from the nanometric single MENPs to micrometric multi-nanofibers structure with dispersed MENPs (poly-ME devices) and ending with the millimetric peripheral nerve, was proposed. MENPs concentration and distribution, device geometry, orientation, and size were systematically varied, and their effects on nerve fiber recruitment were quantified across myelinated and unmyelinated fiber types at different distances from the stimulating source. RESULTS AND DISCUSSION: Our results evidenced how higher MENPs concentrations, higher stimulating source size and lower distance enhance the modulation capability of the poly-ME devices, with differences in stimulation capability between myelinated and unmyelinated nerve fibers. The solution of embedding MENPs in conductive polymer nanofibers enables higher MENPs loading than direct tissue injection, improves positional stability, and preserves biocompatibility, thereby offering a potential alternative for the engineering of MENPs as neuromodulation interfaces.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Nanofiber‐Based Electroactive Interfaces Enabling Coordinated Neuromodulation and Peripheral Nerve RegenerationAdvanced Materials · 2026
- Magnetoelectric Nanoparticles Enable Modulation of Cortical Networks by Low‐Intensity Static Magnetic Fields In VitroAdvanced Science · 2026
- A Multi-Scale Computational Framework for Evaluating Wireless Neuromodulation Feasibility via Magnetoelectric Nanoparticles.Advanced science (Weinheim, Baden-Wurttemberg, Germany) · 2026
- Energy-Efficient Peripheral Magnetic Stimulation via Pulse-Shape Optimization using a Dual-Coil Architecture with Monopolar-Like Nerve-Axis Field.IEEE transactions on neural systems and rehabilitation engineering : a publication of the IEEE Engineering in Medicine and Biology Society · 2026
- Tissue-adaptive bioelectronic fibers with temperature-induced self-tightening enable ultrastable neural interfaceNature Communications · 2026
- Wireless Brain‐Computer Interface With MagnetoElectric NanoParticles: Recording of Neural ActivityAdvanced Theory and Simulations · 2026