High-frequency Bipolar Asymmetric Pulses Enhance Tissue Ablation Efficacy with Low Neuromuscular Stimulation.
Sizhe Xiang, Yancheng Wang, Lisheng Zhao, Chenguo Yao, Shoulong Dong
Key points
From the paper's conclusion- The results demonstrated that, compared with H-FIRE, H-FBAP increased the ablation area by up to 88.77%, while its muscle contraction intensity was only 19.8% of that of IRE at the same amplitude.
Abstract
As a novel tumor ablation technology, irreversible electroporation (IRE) has gradually become one of the important research directions in the field of tumor therapy, owing to its advantages of non-thermal effect and tissue selectivity. However, its clinical application is limited by the key issue of muscle contractions induced by neural stimulation triggered by pulsed electric fields (PEF). This issue has a significant impact on patients' intraoperative treatment and postoperative rehabilita-tion. Although the newly proposed high-frequency irreversible electroporation (H-FIRE) can alleviate neuromuscular stimula-tion, its ablation efficacy is significantly reduced due to the constraint of the "bipolar cancellation" effect. Existing pulse optimization strategies struggle to balance the dual requirements of neuromuscular stimulation inhibition and ablation efficiency improvement. To address this core contradiction, this study proposes a high-frequency bipolar asymmetric pulse (H-FBAP) protocol. This protocol not only retains H-FIRE's advantage of low neuromuscular stimulation but also effectively suppresses the bipolar cancellation effect by optimizing the asymmetric configuration of pulses, thus enhancing ablation efficacy. Meanwhile, this study summarizes the conditions for H-FBAP to inhibit neural excitation and the rules for improving ablation efficiency. The outcomes regarding ablation efficacy and muscle contraction were validated via simulation, in vitro experiments, and animal studies. The results demonstrated that, compared with H-FIRE, H-FBAP increased the ablation area by up to 88.77%, while its muscle contraction intensity was only 19.8% of that of IRE at the same amplitude. This study provides a novel technical solution for addressing the core bottleneck in PEF ablation technology.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Pulse Field Ablation: From Bench to Blueprint: Mechanistic Insights and Device Engineering and Early Clinical Translation.Circulation. Arrhythmia and electrophysiology · 2026
- Clinical Advantages and Pitfalls of Pulsed Field Ablation: How to Prevent ALH Syndrome.Journal of arrhythmia · 2026
- Thermal preconditioning of targeted tissue enhances pulsed field ablation lesionsJournal of Interventional Cardiac Electrophysiology · 2026
- Asymmetric biphasic pulsed field ablation of paroxysmal atrial fibrillation: clinical assessment.Frontiers in cardiovascular medicine · 2026
- Pulsed field ablation versus conventional thermal ablation for atrial fibrillation: a systematic review and meta-analysisEuropean Heart Journal · 2026
- Bipolar needle electrodes for electroporation-based treatments in soft and bone tissues: technical design and implementation.La Radiologia medica · 2026