Thermal preconditioning of targeted tissue enhances pulsed field ablation lesions
Daniel Campos-Villarreal, Maxwell Coll, Logan McClennen, Chieko Imamura, Camila Ciurlo Aragón, Andres F Miranda-Arboleda, Jorge E Romero, Nathaniel Steiger,
Brigham and Women's Hospital
Key points
From the paper's conclusion- These findings suggest that tissue temperature is a tunable factor influencing PFA efficacy.
Abstract
INTRODUCTION: Pulsed field ablation (PFA) is a cardiac ablation modality that induces cell death via irreversible electroporation (IRE) rather than thermal injury. Intrinsic biophysical tissue properties - including temperature and conductivity- can influence PFA lesion formation, however, this phenomenon is not thoroughly characterized. OBJECTIVE: To evaluate the impact of thermal preconditioning on PFA lesion characteristics using a controlled in vitro model. METHODS: Twenty-millimeter-thick Solanum tuberosum potato slabs were assigned to one of five preconditioning groups: cold (7-10 °C), control (21.5-23 °C), room-temperature soaked (immersed in water at 21.5-23 °C), dry-heated (40 °C), and wet-heated (40 °C water-soaked). Pre-ablation surface temperature was recorded. A monophasic unipolar PFA application was delivered (1750 V, 20 µs pulses, 3 pulses, 1-second interval) using a 3.5 mm electrode catheter. Lesions were stained with 2,3,5-triphenyltetrazolium chloride (TTC) and measured. RESULTS: Pre-heating the tissue to 40 °C produced markedly larger lesions than control. Dry-heated samples produced a mean lesion diameter of 34.1 ± 1.7 mm and wet-heated 35.0 ± 1.9 mm, vs. 25.3 ± 2.0 mm in controls (p < 0.001). Lesion depth was also greater in heated tissue (wet-heated 12.23 ± 0.76 mm vs. control 10.7 ± 1.3 mm, p = 0.017). In contrast, cooling the tissue to < 10 °C reduced lesion diameter (23.1 ± 2.3 mm) and depth (9.3 ± 1.0 mm), significantly smaller than controls (both p < 0.01). Room-temperature soaking (RT-soaked) alone did not produce significant change in lesion dimensions. CONCLUSIONS: In this in vitro model, thermal preconditioning significantly alters PFA lesion formation. Increasing the tissue temperature prior to ablation markedly increases lesion diameter and depth, possibly by lowering the electroporation threshold. These findings suggest that tissue temperature is a tunable factor influencing PFA efficacy.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Controlling pulsed field ablation efficacy during ablation procedures.Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology · 2026
- Assessment of temperature dynamics in pulsed field ablation with a variable-loop circular catheter: a comparative analysis of waveform configurations and irrigation rates in specimens of bovine ventricular myocardiumEP Europace · 2025
- Variability in Tissue Interface Temperature During Pulse Field Atrial Ablation: Implications for Real-Time Contact AssessmentCirculation Arrhythmia and Electrophysiology · 2026
- Comparative assessment of the thermal footprints of pulse field ablation systemsEuropean Heart Journal · 2026
- Clinical Advantages and Pitfalls of Pulsed Field Ablation: How to Prevent ALH Syndrome.Journal of arrhythmia · 2026
- Development and validation of a contact force-independent index for predicting pulsed field ablation lesion depthEuropean Heart Journal · 2026