Evaluation of a Novel Hands-Free Cheek Holder for Oscillometry Measurements.
C Veneroni, D Bizzotto, P Pompilio, G Imeri, A Licini, R Dellaca, F Di Marco, M Cottini
Dipartimento di ElettronicaInformazione e Bioingegneria, Politecnico di Milano Milan 20133 Italy. · Restech SrlMilan20124 Italy.
Key points
From the paper's conclusion- The novel cheek holder produced impedance measurements comparable to manual support, simplifying oscillometry and potentially improving its reproducibility.
Abstract
OBJECTIVE: Accurate oscillometry requires minimizing artifacts, particularly those from cheek vibrations, which introduce a parallel (shunt) impedance pathway that distorts measurements. Standard practice includes using a nose clip and manually supporting the cheeks to reduce this effect by stiffening the tissue. This study evaluated the performance of a novel, equipment-integrated cheek holder designed to stabilize the cheeks and improve measurement accuracy without manual support, by comparing it to the current gold-standard (operator-supported cheeks). METHODS: Respiratory resistance (Rrs) and reactance (Xrs) at 5 Hz were measured using a handheld oscillometer (FIRST, Restech srl) with the cheek holder and with manual support, in randomized order. Each test was performed in triplicate. Agreement was assessed using Deming regression. Equivalence was defined as the 95% confidence interval (CI) of the regression line falling within the known short-term repeatability: 17% (Rrs) and 37% (Xrs) for adults; 27% and 40% for children, respectively. RESULTS: 55 subjects (age 5-75 years; height 104-185 cm; 56% females) were included for a broad representation of facial characteristics, with lip length ranging from 36 to 67 mm and bigonial breadth from 80 to 155 mm. The bias between methods was not clinically significant for Rrs or Xrs: the 95% confidence intervals (CI) from the Deming regression fell within the expected short-term variability for all evaluated parameters, indicating consistency between the two methods. CONCLUSION: The novel cheek holder produced impedance measurements comparable to manual support, simplifying oscillometry and potentially improving its reproducibility.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparison of two respiratory oscillometers in patients with a broad range of lung function patterns.Frontiers in allergy · 2026
- Assessing pulmonary function in 3-year-old children using respiratory oscillometry: a validation study in rural Guatemala.ERJ open research · 2026
- Clinical validation of the MED-58 upper-arm automated oscillometric blood pressure monitor with fan-shaped and rigid cuffs for home and clinical use in adolescents and adults, according to International Organization for Standardization 81060-2:2018 (Amd.1:2020; Amd.2:2024): a single-center prospective study.Blood pressure monitoring · 2026
- Dental Occlusion Measurement: A Methodological Evaluation to Improve Clinical Gnathology.Sensors (Basel, Switzerland) · 2026
- An Oscillometric Readout Assessing the Degree of Airway Smooth Muscle Activation In Vivo.Annals of biomedical engineering · 2026
- Intraoperative non-contact three-dimensional measurement of the facial recess using a 4K3D exoscope during cochlear implantation.European archives of oto-rhino-laryngology : official journal of the European Federation of Oto-Rhino-Laryngological Societies (EUFOS) : affiliated with the German Society for Oto-Rhino-Laryngology - Head and Neck Surgery · 2026