Assessing pulmonary function in 3-year-old children using respiratory oscillometry: a validation study in rural Guatemala.
Laura M Grajeda, Keyla Castellanos, Tatiana Petrovick, Anaité Díaz-Artiga, María R López, Albino Barraza-Villarreal, Lisa M Thompson, Christina M Eckhardt,
Global Health Institute, Department of Epidemiology and Biostatistics, University of Georgia, Athens, GA, USA. · Centro de Estudios en Salud, Universidad del Valle de Guatemala, Guatemala, Guatemala.
Key points
From the paper's conclusion- We collected oscillometry results from 537 children. Height explained 5-15% of the variability in R7, X7 and AX, but none in R7-R19. The correlation between standard and manually cleaned tests ranged from 0.66 to 0.97, depending on the parameter. Of the first tests, between-day ICCs using the standard method for R7, R7-R19, X7 and AX were 0.68, 0.45, 0.60 and 0.71, respectively, and the best between-day ICCs using an alternative method were 0.75, 0.57, 0.69 and 0.74, respectively.
Abstract
BACKGROUND: Respiratory oscillometry can inform early-life lung function, but evidence supporting its use in Guatemala is limited. We assessed the validity and reliability of respiratory oscillometry in 3-year-old Guatemalan children and whether these improve with quality control (QC). METHODS: During home visits, we measured respiratory oscillometry resistance (R) and reactance (X) at 7 Hz (R7 and X7), the frequency dependence of resistance (R7-R19) and the area of reactance (AX) in 666 children participating in the Household Air Pollution Intervention Network trial. Standard measurement QC removed artefacts using commercial software and selected the first three measurements with an R7 coefficient of variation of ≤15%. Alternative investigators' methods varied in QC procedures. We evaluated validity as the proportion of variability in tests predicted by a child's height (R2) and, in a subset of 50 children, as agreement with test results after manual QC by a paediatric pulmonologist. Reliability was assessed by the intraclass correlation coefficient (ICC) of repeated tests. RESULTS: We collected oscillometry results from 537 children. Height explained 5-15% of the variability in R7, X7 and AX, but none in R7-R19. The correlation between standard and manually cleaned tests ranged from 0.66 to 0.97, depending on the parameter. Of the first tests, between-day ICCs using the standard method for R7, R7-R19, X7 and AX were 0.68, 0.45, 0.60 and 0.71, respectively, and the best between-day ICCs using an alternative method were 0.75, 0.57, 0.69 and 0.74, respectively. DISCUSSION: We demonstrated good validity and reliability of respiratory oscillometry measurements collected at home from preschool-aged Guatemalan children. Investigators' automated QC methodology performed equally to or better than standard methodology.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Use of respiratory oscillometry in infants: a scoping reviewPediatric Research · 2026
- Comparison of two respiratory oscillometers in patients with a broad range of lung function patterns.Frontiers in allergy · 2026
- Respiratory Impedance Prediction in Children and Adolescents: A Meta-Modelling Approach.Annals of biomedical engineering · 2026
- Respiratory oscillometry reference values in Polish children and adolescents: a nationwide multicenter cross-sectional study (OSCILLONiCA-PL-Kids) - Study Protocol.Advances in medical sciences · 2026
- Machine Learning-Based Prediction of Pulmonary Function Test Outcomes Using Oscillometry in Patients with COPD.Sensors (Basel, Switzerland) · 2026
- Performance of automated respiratory rate counters for respiratory rate assessment to support pneumonia detection among children under five years of age in low- and middle-income settings: a systematic review of qualitative and quantitative evidence.Journal of global health · 2026