Physics in medicine and biology

Scatter correction for multi-source linear static CT via a geometry-aware parameter regression network with differentiable forward modeling.

Yankun Wang, Yule Li, Junhao Wu, Deyang Kong, Haomin Hou, Hao Wang, Zhaoying Bian, Limin Li,

Southern Medical University, School of Biomedical Engineering, Guangzhou, 510515, China. · Xi'an Jiaotong University, School of Mathematics and Statistics, Xi'an, 710049, China.

Key points

From the paper's conclusion
  • To address this challenge, we propose a scatter correction framework that integrates a geometry-aware analytical kernel with a parameter regression network (GAPR-Net).

Abstract

OBJECTIVE: Multi-source linear static CT (SLCT) offers high temporal resolution and a compact form factor by eliminating mechanical rotation. However, lateral offsets of peripheral X-ray beams break the rotational symmetry assumed by conventional scatter correction methods, resulting in location-dependent and asymmetric scatter point spread functions (PSFs). APPROACH: To address this challenge, we propose a scatter correction framework that integrates a geometry-aware analytical kernel with a parameter regression network (GAPR-Net). An explicit correction term is incorporated into the conventional cone-beam scatter kernel to model asymmetric PSF distortion induced by lateral source offsets. GAPR-Net learns scatter kernel hyperparameters associated with source geometry and object thickness from Monte Carlo pencil-beam data, while a differentiable forward model analytically estimates scatter and enables stable optimization under a geometric consistency loss. MAIN RESULTS: Extensive simulation experiments were performed using water, electron-density, and ICRP phantoms, together with a DICOM-derived head phantom reconstructed using the derived three-dimensional SLCT reconstruction algorithm. Compared with SKS, DSE, ResNet, and a static-CT-specific 1D U-Net, GAPR-Net achieved consistent improvements. For the electron-density and ICRP phantoms, GAPR-Net achieved scatter-estimation mean absolute percentage errors of 4.43% and 5.31%, respectively. In the reconstructed images, it achieved a peak signal-to-noise ratio of 29.51 dB for the electron-density phantom and reduced the mean absolute percentage error of the ICRP thorax phantom from 14.69% to 6.07%. Head-phantom experiments at 80 kVp, 100 kVp, and 120 kVp further demonstrated stable performance across different X-ray spectra. A physical testbed experiment was conducted to evaluate the performance of the proposed GAPR-Net under real conditions. SIGNIFICANCE: GAPR-Net provides a physically interpretable and computationally efficient approach for correcting non-stationary asymmetric scatter artifacts in SLCT.&#xD.

Read the paper