Journal of thoracic imaging

Optimizing AI-Derived CAC Volume Thresholds on Non-Gated Chest CT for Detection of Actionable Coronary Artery Calcium: A Validation Against Agatston Scoring and Radiologist Reporting.

Maya Wiessman, James Wyatt Miller, Ann Marie Navar, Parag Joshi, Yin Xi, Xinhui Duan, Travis Browning, Arzu Canan,

Department of Internal Medicine, Division of Cardiology. · School of Medicine.

Observational · n=39

Key points

From the paper's conclusion
  • AI-derived CAC volume from non-gated chest CTs correlates strongly with reference CAC scores, showing high sensitivity at actionable thresholds. This supports potential opportunistic screening, though impact on outcomes requires study.

Abstract

PURPOSE: Coronary artery calcium (CAC) visually estimated on non-electrocardiogram (ECG)-gated chest CTs is reported inconsistently, limiting identification of high-risk individuals. Although automated AI-derived CAC scoring offers the potential to standardize reporting and improve reliability, this method warrants rigorous validation. MATERIALS AND METHODS: Three hundred thirty-two patients (59% female, median age 62) who underwent a non-gated chest CT within 1-year of a dedicated cardiac CT with Agatston's CAC score (CACscore) were identified between 2016 and 2022. AI-Rad Companion (Siemens) was used to compute CAC volume (AI-CACvol). AI-CACvol performance was compared with radiologists' qualitative CAC reports (H-CACqual) using optimized thresholds for identifying actionable CAC. Model performance was further validated in a postimplementation population. RESULTS: Median AI-CACvol was 20.3 mm³ (IQR 0-188.65), strongly correlated with CACscore (Spearman ρ=0.91, 95% CI: 0.88-0.94); median 31.55 Agatston Units (AU), IQR: 0 to 250). Agreement with CACscore was higher for AI-CACvol versus H-CACqual (weighted kappa 0.88, 95% CI: 0.84, 0.91; vs. 0.85, 95% CI: 0.82, 0.88; P=0.022). At a cutoff of 70 mm³, AI-CACvol demonstrated superior sensitivity compared with H-CACqual for detecting CAC≥100 AU, with greater sensitivity (95% vs. 75%) and comparable specificity (92% vs. 94%). In a validation cohort of 39 patients (64% female, median age 67 y), median CACscore was 43 AU (IQR 0-215) and median AI-CACvol 29.7 mm³ (IQR: 0 to 219.4). The same 70 mm³ cutoff achieved 100% sensitivity and 88% specificity for CAC≥100 AU (n=13). CONCLUSION: AI-derived CAC volume from non-gated chest CTs correlates strongly with reference CAC scores, showing high sensitivity at actionable thresholds. This supports potential opportunistic screening, though impact on outcomes requires study.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers