Artificial Intelligence Detection of Aortic Valve Calcification on Routine Noncontrast Chest CT: A Real-World Cohort.
Ofir Koren, Vivek Patel, Orit Wimpfheimer, Alexander Kukuev, Parker Chu, Harsh Jain, Kazuki Suruga, Hidemasa Shitan,
Smidt Heart Institute, Cedars-Sinai Medical Center, Los Angeles, California, USA. Electronic address: Ofir.Koren@csmc.edu. · Smidt Heart Institute, Cedars-Sinai Medical Center, Los Angeles, California, USA.
Key points
From the paper's conclusion- AVC was common but infrequently documented on routine chest CT. AI detected most cases among evaluable scans; 11.3% were nonevaluable, yielding 89.9% end-to-end sensitivity. Prospective studies should assess clinical pathways.
Abstract
BACKGROUND: Routine noncontrast chest computed tomography (CT) may reveal aortic valve calcification (AVC) outside aortic stenosis evaluation. Its prevalence, burden, and documentation remain uncertain. Artificial intelligence (AI) may enable systematic detection. OBJECTIVES: The study aims to evaluate AVC prevalence and burden, clinical correlates, radiology documentation, and AI-assisted case finding. METHODS: This retrospective single-center study screened 511 consecutive noncontrast, non-electrocardiographically gated chest CT examinations in adults aged ≥50 years. After 15 protocol exclusions and 56 software-designated nonevaluable examinations, 440 underwent quantification by 2 blinded readers. AVC was defined as >10 AU using a 130-Hounsfield-unit threshold adjusted for blood-pool attenuation and image noise when needed. Locked HealthAVC V1.0 was a secondary comparator. RESULTS: Manual quantification identified AVC in 172 of 440 scans (39.1%; 95% CI: 34.6%-43.7%); 97 of 440 (22.0%) had ≥100 AU. Inter-reader agreement at >10 AU was 99.5% (kappa = 0.990). Age (adjusted OR: 1.92 per 5 years; 95% CI: 1.63-2.26) and male sex (adjusted OR: 2.09; 95% CI: 1.32-3.30) were associated with AVC. Reports documented 27 of 172 positive scans (15.7%; 95% CI: 11.0%-21.9%). AI detected 169 of 172 manually confirmed cases (98.3%) among 440 evaluable scans, including 142 of 145 undocumented cases. Of 496 submitted scans, 56 (11.3%) were nonevaluable. End-to-end sensitivity counting positive rejected scans as missed was 89.9%. Among 86 AI-positive patients with echocardiography, 10 (11.6%) had mild-or-greater aortic stenosis. CONCLUSIONS: AVC was common but infrequently documented on routine chest CT. AI detected most cases among evaluable scans; 11.3% were nonevaluable, yielding 89.9% end-to-end sensitivity. Prospective studies should assess clinical pathways.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Automated Aortic Valve Calcification Scoring: Multicenter External Validation of a Deep Learning Algorithm.JACC. Advances · 2026
- Validation of Artificial Intelligence-based Non-gated Chest CT for Coronary Artery Calcium Scoring Across Multiple CT Scanners.Journal of imaging informatics in medicine · 2026
- Optimizing AI-Derived CAC Volume Thresholds on Non-Gated Chest CT for Detection of Actionable Coronary Artery Calcium: A Validation Against Agatston Scoring and Radiologist Reporting.Journal of thoracic imaging · 2026
- Fully automatic coronary artery calcium score using CT and artificial intelligence: external validation on a danish populationEuropean Heart Journal · 2026
- Opportunistic Detection of Coronary Artery Calcium on Noncardiac Chest Computed Tomography: An Emerging Tool for Cardiovascular Disease Prevention: A Scientific Statement From the American Heart AssociationCirculation · 2025
- Radiomics and artificial intelligence for coronary calcium scoring: insights from the CAC CV-PREVITAL studyEuropean Heart Journal · 2026