Diagnostic Feasibility and Physical Assessment of Lung CT Images Reconstructed From Ultra-Low-Dose Positioning Scan for Computer-Aided Nodule Detection.

Hiroyuki Morisaka, Kaori Aoyagi, Hiroaki Watanabe, Tomomi Yato, Junichi Sato, Shintaro Ichikawa

Department of Diagnostic Radiology, University of Yamanashi Faculty of Medicine. · Department of Radiological Technology, University of Yamanashi Hospital, Chuo, Yamanashi, Japan.

Observational · n=160

Key points

From the paper's conclusion
  • Thinner-slice reconstruction improved CAD sensitivity on volumetric ULD positioning scans; however, noninferiority to standard-dose CT was not demonstrated. Further optimization of image quality and CAD performance is warranted.

Abstract

OBJECTIVE: To evaluate the diagnostic feasibility of computer-aided detection (CAD) of pulmonary nodules on CT images reconstructed from volumetric ultra-low-dose (ULD) positioning scans compared with standard-dose CT. METHODS: In this retrospective study, 160 adults with at least 1 pulmonary nodule underwent volumetric ULD positioning scans and standard-dose CT. One target nodule per patient was analyzed. CAD sensitivity was evaluated on 5-mm volumetric ULD positioning scans and 5-mm standard-dose CT according to nodule size, morphology, and malignancy. An additional analysis was performed using 1-mm volumetric ULD positioning scans, with 1-mm standard-dose CT serving as the reference standard. Physical image characteristics were assessed using phantoms. RESULTS: A total of 160 nodules (median size: 5 mm) were analyzed. CAD sensitivity was 71.3% (114/160) for 5-mm volumetric ULD positioning scans and 85.0% (136/160) for 5-mm standard-dose CT (P<0.01), with reduced sensitivity most apparent for small nodules. With 1-mm volumetric ULD positioning scans, sensitivity improved to 88.8% (142/160). However, noninferiority to reference 1-mm standard-dose CT was not demonstrated using a 15-percentage-point margin (P=0.084). The median estimated effective dose was 0.18 mSv for volumetric ULD positioning scans and 4.67 mSv for standard-dose CT. Phantom assessment showed lower spatial resolution and increased low-frequency noise for volumetric ULD positioning scans. CONCLUSION: Thinner-slice reconstruction improved CAD sensitivity on volumetric ULD positioning scans; however, noninferiority to standard-dose CT was not demonstrated. Further optimization of image quality and CAD performance is warranted.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers