Hidden Risks of Multimodal Neuromodulation: Suspected Intrathecal Baclofen Pump Motor Stall Induced by a Vagus Nerve Stimulator Magnet.

Priya Sharma, Brigid Deck, Gregor J Dudiak, Weibin Shi, Thiru Annaswamy

原始摘要(英文原文)· Original abstract
The increasing use of implantable neuromodulation devices in patients with chronic post-stroke sequelae raises the possibility of unintended device-device interactions. We report a case of recurrent intrathecal baclofen pump motor stalls in a patient with longstanding spasticity following an intracranial hemorrhage, ultimately attributed to magnetic interference from a recently implanted vagus nerve stimulator (VNS) activator magnet. Recognition of this interaction led to prompt resolution and highlights the importance of interdisciplinary awareness and patient education in individuals with multiple implanted neuromodulation systems.