Magnetic interference and programmable valves: navigating hidden risks in everyday life and clinical settings.

Arianna Barbotti, Camilla de Laurentis, Edoardo Porto, Federico Di Rocco, Pierre-Aurélien Beuriat, Matthieu Vinchon, Carmine Mottolese, Alexandru Szathmari

Neurosurgery Department, Fondazione IRCCS Istituto Neurologico Nazionale "C. Besta", Milan, Italy. · Neurochirurgie Pédiatrique Hôpital Femme Mère Enfant, Hospices Civils de Lyon, Université de Lyon 1, Lyon, France.

Review

Key points

From the paper's conclusion
  • We mapped evidence on magnetic interference from medical and everyday sources.

Abstract

Programmable shunt valves, widely used in hydrocephalus, especially in children, allow non-invasive pressure adjustment, but their magnetic rotors make them vulnerable to unintended setting changes. We mapped evidence on magnetic interference from medical and everyday sources. We conducted a scoping review (PRISMA-ScR) of PubMed, Embase and Scopus (to 20 September 2026), extracting design, valve, magnetic source, effect and reversibility, with descriptive quality appraisal. Of 674 records, 51 studies were included: 31 ex vivo or in vitro, 10 clinical series, 9 case reports and 1 of unstated design. Sources included magnetic resonance imaging (MRI), other medical and hearing devices, and everyday objects (magnetic toys, smartphones, headphones, tablets). Quality, appraised in 47 full texts, was variable. In two large clinical series, settings changed after 26.7% and 56.7% of MRI examinations, mainly in Strata but also in Polaris (31%), Certas Plus (12%) and proGAV (10%), never in the few Strata MR or proGAV 2.0; a randomized trial found none in Certas Plus. Ex vivo, locking valves resisted 3 T; at 7 T, outside their labeling, Certas Plus and proGAV 2.0 lost reprogrammability. Everyday devices altered most valves, mainly at contact, except proGAV models and M.blue Plus; several bench studies found no change. Susceptibility varies by model and appears highest in valves without a mechanical lock. MRI is the best-documented clinically relevant source of setting changes; settings should be verified afterwards in all valves. Given limited evidence, distance from magnets and caregiver education, particularly for children, are precautions rather than proven safety limits.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers