Comprehensive acceptance testing and quality control of Single Photon Emission Computed Tomography/Computed Tomography systems.
Noor Ul Huda Abbasi, Hannan Younis, Saeed Ur Rehman, Burhan Zia Abbasi, Roya Feroze, Khurram Mehboob, Muhammad Ajaz
Radiation Physics Lab, Department of Physics, COMSATS University Islamabad, . · Department of Radiology, NORI Cancer Hospital, Islamabad, Pakistan, .
Key points
From the paper's conclusion- The SPECT/CT systems at NORI Hospital Islamabad met established NEMA standards for image quality. Acceptance testing validated detector performance and highlighted minor areas for improvement, such as periodic sensitivity calibration. Implementation of a structured quality control program ensures consistent equipment performance, optimal image quality, improved diagnostic accuracy, and superior patient care. Routine adherence to international standards is essential for maintaining high-quality nuclear medicine imaging practices.
Abstract
BACKGROUND: Single photon emission computed tomography/computed tomography (SPECT/CT) systems play a crucial role in nuclear medicine for accurate diagnosis and treatment planning. Ensuring high image quality through rigorous acceptance testing and quality control is essential to maintain diagnostic reliability and patient safety. OBJECTIVE: This study evaluated the performance of SPECT/CT systems at NORI Hospital Islamabad against National Electrical Manufacturers Association (NEMA) and International Atomic Energy Agency standards and developed a structured quality control program based on acceptance testing results. METHODS: Acceptance tests were carried out using Tc-99m, I-131, and Co-57 radioactive sources, in conjunction with quadrant bar and Jaszczak phantoms. Parameters assessed included spatial resolution, intrinsic and extrinsic uniformity, energy resolution, sensitivity, center of rotation (COR), detector shielding, background counts, leakage, CT number accuracy, and nuclear medicine/CT registration. Data were compared with manufacturer specifications and NEMA standards to evaluate system performance and inform quality control strategies. RESULTS: Both detectors exhibited intrinsic and extrinsic uniformity within NEMA limits, ensuring consistent image quality. Spatial resolution testing confirmed sharp and precise imaging, while energy resolution was within manufacturer-specified values. Sensitivity measurements showed minor differences between detectors but remained within acceptable ranges. COR values were precise (<±0.55 mm), confirming proper mechanical and software alignment. CT number accuracy and nuclear medicine/CT registration tests confirmed precise anatomical-functional image overlay. Detector shielding and leakage tests showed minimal interference, supporting reliable imaging. CONCLUSION: The SPECT/CT systems at NORI Hospital Islamabad met established NEMA standards for image quality. Acceptance testing validated detector performance and highlighted minor areas for improvement, such as periodic sensitivity calibration. Implementation of a structured quality control program ensures consistent equipment performance, optimal image quality, improved diagnostic accuracy, and superior patient care. Routine adherence to international standards is essential for maintaining high-quality nuclear medicine imaging practices.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Development and validation of a modular multi-purpose phantom for CT, PET/CT, and SPECT/CT quality control.Physical and engineering sciences in medicine · 2026
- Impact of CT parameter reduction on quantitative SPECT/CT assessment in 99mTc imaging: A phantom study.Radiography (London, England : 1995) · 2026
- Systematic evaluation of energy window and reconstruction parameters for quantitative lutetium‐177 SPECT imaging on a CZT‐based starguide SPECT/CT systemJournal of Applied Clinical Medical Physics · 2026
- Monte Carlo simulation and experimental validation of SPECT systems via energy spectrum decomposition and image profile analysis.EJNMMI physics · 2026
- From static to dynamic SPECT/CT: A systematic review of clinical applications and perspectives.Seminars in nuclear medicine · 2026
- Development of a multifunctional anthropomorphic phantom for performance evaluation in nuclear medicine imaging: preliminary PET/CT validation.EJNMMI physics · 2026