Planning evaluation of 3 photon techniques for hippocampus and scalp sparing WBRT with simultaneous integrated boost in multiple metastases.
Huaqu Zeng, Yangxing Lin, Zhen Li, Zongyou Chen, Shukui Tang, Qifu Lin, Minzhi Zhong, Ming Chen,
Radiotherapy Center, Gaozhou People's Hospital, Gaozhou, Guangdong, China. · Department of Oncology Radiotherapy, Zhongshan City People's Hospital, Zhongshan, Guangdong, China.
Key points
From the paper's conclusion- This study compared the dosimetric and radiobiological parameters of 3 photon beam techniques - 6 megavoltage (MV) flattening filter-free (6XFFF), 6 MV flattened (6XFF), and 10 MV flattened (10XFF) - in hippocampus and scalp sparing whole-brain radiotherapy with simultaneous integrated boost for multiple brain metastases.
Abstract
This study compared the dosimetric and radiobiological parameters of 3 photon beam techniques - 6 megavoltage (MV) flattening filter-free (6XFFF), 6 MV flattened (6XFF), and 10 MV flattened (10XFF) - in hippocampus and scalp sparing whole-brain radiotherapy with simultaneous integrated boost for multiple brain metastases. Twenty-two patients with multiple brain metastases underwent volumetric modulated arc therapy planning using 6XFFF, 6XFF, and 10XFF beams. All plans adhered to Radiation Therapy Oncology Group 0933 constraints for hippocampal avoidance and scalp sparing. Target coverage was evaluated using parameters such as V50 ≥ 95% for planning gross tumor volume and V30 ≥ 90% for planning target volume of the brain. Dosimetric parameters for organs at risk, including the hippocampus and scalp, were analyzed. Radiobiological evaluation included tumor control probability and normal tissue complication probability. Delivery efficiency was assessed using monitor units and control points. All 3 techniques achieved comparable target coverage with no significant differences in dose distribution (P > .05). The 10XFF plan tended to reduce scalp doses (mean Dmax = 31.18 ± 5.4 Gy; Dmean = 15.30 ± 1.23 Gy) compared to 6XFF and 6XFFF, though the differences were not statistically significant (P > .05). Normal tissue complication probability for the hippocampus and tumor control probability for targets were similar across techniques. Delivery efficiency was superior for flattened beams, with the 10XFF plan requiring significantly fewer monitor units (1277.81 ± 178.87) and control points (367.40 ± 19.53) than the 6XFFF plan (P < .001). All 3 techniques provided clinically acceptable target coverage while effectively sparing the hippocampus and scalp. The 10 MV flattened photon beam showed a trend toward reduced scalp irradiation and improved delivery efficiency, suggesting its potential utility when scalp preservation or workflow efficiency is prioritized. Further prospective studies are warranted to validate these findings.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparative Dosimetric Evaluation of Tomotherapy with a 1-cm Field Width versus Volumetric-modulated Arc Therapy for Whole-brain Radiotherapy with Hippocampal Avoidance.Journal of medical physics · 2026
- A Dosimetric and Biological Comparison of Volumetric Modulated Arc Therapy and Adaptive Aperture Proton Therapy for Hippocampal-Sparing Whole-Brain Radiotherapy.Technology in cancer research & treatment · 2026
- Radiotherapy Modality and Primary Tumor Type as Determinants of Outcomes in Brain Metastases: A Retrospective Study of 734 Patients.Medicina (Kaunas, Lithuania) · 2026
- Dosimetric Comparison of Contemporary Stereotactic Radiotherapy Platforms for Brain Metastases: Towards Personalized Technology Selection.Technology in cancer research & treatment · 2026
- Dosimetric comparison of automated noncoplanar volumetric-modulated arc therapy and intensity-modulated proton therapy for total scalp irradiation in angiosarcoma of the scalp.The British journal of radiology · 2026
- Updated dosimetric comparison of HDR interventional radiotherapy including 60Co sources versus VMAT and electron beam therapy for NMSC.Radiation and environmental biophysics · 2026