Effect of heparin on immunological heparin-induced thrombocytopenia-IgG chemiluminescence testing.
Ronny Vong, Leonardo Pasalic, Dea Donikian, Timothy Brighton, Emmanuel J Favaloro
Department of Haematology, Institute of Clinical Pathology and Medical Research (ICPMR), NSW Health Pathology. · Coagulation and Special Haemostasis, Department of Haematology, New South Wales Health Pathology (NSWHP) East, Prince of Wales Hospital, Randwick.
Key points
From the paper's conclusion- High-dose heparin, used for inhibiting serotonin release in SRA, had a variable effect on HIT-IgG testing, generally reducing detected HIT-IgG levels in high positive samples, and mostly reducing detected HIT-IgG levels in low positive samples.
Abstract
Heparin-induced thrombocytopenia (HIT) requires clinical evaluation and laboratory testing using immunological assays with confirmation by functional assays such as serotonin release assay (SRA). The HIT-IgG chemiluminescence immunoassay (CLIA) is a rapid immunological assay; we aimed to assess the effect of heparin on this assay. The effect of low-dose (0.1 U/ml) and high-dose (100 U/ml) heparin on the CLIA HIT-IgG assay was evaluated, these concentrations aligning to those used for SRA. A cohort of available pretested patient material (plasma or serum) comprised samples negative for HIT-IgG (i.e., <1.0 U/ml) (n = 22), or positive for HIT-IgG (i.e., ≥1.0 U/ml) (n = 25), with the latter split into 'low positive' (1.0-<10 U/ml; n = 16) and 'high positive' (≥10 U/ml; n = 9). Low-dose heparin, used for initiating serotonin release in SRA, had no appreciable effect on the CLIA HIT-IgG assay for any HIT-IgG group. High-dose heparin, used for inhibiting serotonin release in SRA, had a variable effect on HIT-IgG testing, generally reducing detected HIT-IgG levels in high positive samples, and mostly reducing detected HIT-IgG levels in low positive samples. However, some samples from negative and low positive groups showed counterintuitive increases in HIT-IgG signal. Heparin levels may influence HIT-IgG test values in various ways. Although high positive samples showed an expected decrease in signal with high-dose heparin, some low positive and negative samples showed unexpected increases in signal. Whether these findings improve identification of pathological HIT, or of variable HIT subtypes, requires further investigation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Performance evaluation of a reflexive algorithm for suspected heparin-induced thrombocytopenia.Research and practice in thrombosis and haemostasis · 2026
- A-230 Establishing an Optimal HIT Optical Density Value: A Six Year StudyClinical Chemistry · 2026
- Ticagrelor-Induced False-Negative Serotonin Release Assay in Heparin-Induced Thrombocytopenia With Stent ThrombosisJACC Case Reports · 2026
- Saline whole-blood impedance aggregometry for detection of heparin-independent platelet-activating anti-platelet factor 4 antibodies.American journal of clinical pathology · 2026
- The hepatic FGFR-ERK-HRG axis regulates heparin-induced thrombocytopenia with thrombosis.Blood · 2026
- A buffer-positive serotonin release assay pattern is associated with increased mortality in HIT: a retrospective cohort study.Research and practice in thrombosis and haemostasis · 2026