Impact of Cannulation on Venous Congestion, Right Ventricular Function, and End-Organ Protection During VV ECMO: A Computational Study.
Youssef El Dsouki, Ignazio Condello, Simona Celi, Jan Coveliers, Paolo Meani, Matteo Rossetti, Gennaro Martucci, Roberto Lorusso
From the Maastricht University Medical Centre (MUMC), Maastricht, the Netherlands. · School of Medicine and Surgery, University of Insubria, Varese, Italy.
Key points
From the paper's conclusion- These findings suggest that cannulation geometry may be an important determinant of venous decongestion, right-heart interaction, and simulated cardio-renal-hepatic protection beyond extracorporeal flow alone.
Abstract
Cannulation strategy during veno-venous extracorporeal membrane oxygenation (VV-ECMO) may influence venous drainage, right ventricular loading, and end-organ perfusion beyond gas exchange. We compared four cannulation strategies using a computational hemodynamic framework: pulmonary dual-lumen, Avalon dual-lumen, femoro-jugular, and femoro-femoral VV-ECMO, with three cannula sizes/configurations for each strategy and 10 virtual adult patients per configuration (120 simulations). Cannulation strategy was strongly associated with residual venous congestion and simulated organ dysfunction. Pulmonary dual-lumen support showed the most favorable size-dependent profile. The 31 Fr configuration achieved the lowest central venous pressure (8.8 ± 1.2 mm Hg), smallest inferior vena cava diameter (17.6 ± 1.6 mm), and greatest improvement in tricuspid annular plane systolic excursion (+5.4 ± 1.0 mm). It was also associated with the lowest creatinine, bilirubin, simulated dialysis risk, and hepatic dysfunction. Femoro-femoral VV-ECMO showed the least favorable profile despite preserved flow capability. These findings suggest that cannulation geometry may be an important determinant of venous decongestion, right-heart interaction, and simulated cardio-renal-hepatic protection beyond extracorporeal flow alone.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Influence of side-hole configuration on recirculation in dual-lumen cannulas for veno-venous extracorporeal membrane oxygenation: a parametric computational fluid dynamics study.Frontiers in bioengineering and biotechnology · 2026
- Cannula tip separation and configuration as determinants of recirculation in venovenous ECMO.The journal of extra-corporeal technology · 2026
- Hemodynamic Effects of Three Different Mechanical Circulatory Support Strategies in Isolated Acute Right Heart Failure.ASAIO journal (American Society for Artificial Internal Organs : 1992) · 2026
- Impact of side hole design of drainage cannulas on flow velocity and predilection sites for low-flow areas in veno-venous ECMO circuits.Intensive care medicine experimental · 2026
- Cannula Pairing Determines Distal Limb Perfusion During Peripheral Veno-Arterial Extracorporeal Membrane Oxygenation: An Experimental Mock-Loop Study.Artificial organs · 2026
- Extracorporeal Membrane Oxygenation Related Axillary Artery Cannulation and Aortic Hemodynamics: Insights From Computational Fluid DynamicsASAIO Journal · 2026