Influence of side-hole configuration on recirculation in dual-lumen cannulas for veno-venous extracorporeal membrane oxygenation: a parametric computational fluid dynamics study.
Chenying Zhu, Ping Ye, Zhaohua Chang
School of Health Science and Engineering, University of Shanghai for Science and Technology, Shanghai, China.
Key points
From the paper's conclusion- Results indicate that the recirculation fraction (RF) is influenced by two geometric factors. First, among six configurations with fixed axial spacing, the peak RF (RF_max) was positively associated with the SVC drainage fraction (R 2 = 0.87, n = 6). Given the limited number of configurations and the narrow range of SVC drainage fractions examined, this association should be considered exploratory. Second, reducing axial spacing increased the time-averaged RF (RF_ave) from 2.73% in the baseline configuration to 6.48%. Circumferential distribution of the IVC side holes was associated with a lower RF_ave (2.10%-2.34%) and with reduced stagnation volume, without increased WSS exposure or impaired washout.
Abstract
INTRODUCTION: Veno-venous extracorporeal membrane oxygenation (VV-ECMO) increasingly utilizes dual-lumen cannulas (DLCs) to facilitate single-site cannulation. However, challenges such as recirculation and elevated wall shear stress (WSS) continue to impede their efficacy. The influence of the geometric configuration of drainage side holes on these hemodynamic parameters remains inadequately characterized. METHODS: In this study, we performed transient, non-Newtonian computational fluid dynamics (CFD) simulations on eight distinct DLC configurations subjected to pulsatile venous inflow. These simulations were conducted within an idealized model of the vena cava and right atrium (RA), developed in accordance with ISO 18193:2021 standards. The study systematically varied the axial spacing between drainage holes and the infusion port (denoted as H1/H2), as well as the circumferential and axial arrangement of side holes in the superior vena cava (SVC) and inferior vena cava (IVC). RESULTS: Results indicate that the recirculation fraction (RF) is influenced by two geometric factors. First, among six configurations with fixed axial spacing, the peak RF (RF_max) was positively associated with the SVC drainage fraction (R 2 = 0.87, n = 6). Given the limited number of configurations and the narrow range of SVC drainage fractions examined, this association should be considered exploratory. Second, reducing axial spacing increased the time-averaged RF (RF_ave) from 2.73% in the baseline configuration to 6.48%. Circumferential distribution of the IVC side holes was associated with a lower RF_ave (2.10%-2.34%) and with reduced stagnation volume, without increased WSS exposure or impaired washout. DISCUSSION: These results suggest that considering circumferential side-hole placement could be a useful and underexamined parameter for DLC design optimization in VV-ECMO within the investigated geometric range.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Impact of side hole design of drainage cannulas on flow velocity and predilection sites for low-flow areas in veno-venous ECMO circuits.Intensive care medicine experimental · 2026
- Impact of Cannulation on Venous Congestion, Right Ventricular Function, and End-Organ Protection During VV ECMO: A Computational Study.ASAIO journal (American Society for Artificial Internal Organs : 1992) · 2026
- Cannula tip separation and configuration as determinants of recirculation in venovenous ECMO.The journal of extra-corporeal technology · 2026
- Hemodynamic optimization of a novel single-puncture dual-lumen coaxial catheter for arteriovenous fistula access: A computational fluid dynamics study.Northern clinics of Istanbul · 2026
- Computational fluid dynamics assessment of a triple-lumen hemodialysis catheter.The journal of vascular access · 2026
- Cannula Pairing Determines Distal Limb Perfusion During Peripheral Veno-Arterial Extracorporeal Membrane Oxygenation: An Experimental Mock-Loop Study.Artificial organs · 2026