Clinically Motivated Redesign of the Pennsylvania State University Child Ventricular Assist Device.
Christopher M Scheib, Eric E Yeager, Lichong Xu, Christopher A Siedlecki, Joshua P Cysyk, Karl O Bohnenberger, Choon-Sik J Jhun
From the Division of Applied Biomedical Engineering, Department of Surgery, College of Medicine, Penn State University, Hershey, Pennsylvania.
Key points
From the paper's conclusion- Building on validated in vitro and chronic in vivo testing of the Penn State University (PSU) Child Ventricular Assist Device (CVAD), the four-bladed prototype rotor (Rev 0.0) was redesigned using established analytical and computational methods to improve hydraulic efficiency while preserving hemodynamic performance and biocompatibility.
Abstract
Building on validated in vitro and chronic in vivo testing of the Penn State University (PSU) Child Ventricular Assist Device (CVAD), the four-bladed prototype rotor (Rev 0.0) was redesigned using established analytical and computational methods to improve hydraulic efficiency while preserving hemodynamic performance and biocompatibility. Applying turbomachinery principles at the nominal operating condition (3 liters/minute [LPM], 8,000 revolutions/minute [RPM]), a five-bladed design (Rev 1.0) was created with an ideal inflow angle (β1 = 41.0°) and outflow angle (β2 = 27.4°). In silico simulations of Rev 1.0 across the prototype's operating range (7,000-9,000 RPM, 1-5 LPM) demonstrated an approximately 40% improvement in hydraulic performance. The Rev 1.0 rotor was then evaluated on a benchtop mock circulatory loop using fresh ovine blood, confirming the in silico gains. The normalized index of hemolysis (NIH), measured at two clinically relevant conditions (1.5 LPM and 3.3 LPM, both at ~80 mm Hg), was 0.052 ± 0.054 and 0.047 ± 0.033 mg/dL, respectively, and high-molecular-weight multimers of von Willebrand factor were preserved within normal to borderline ranges. By achieving target flows and pressures at approximately 1,000 RPM lower speeds, the redesign reduces power requirements and enables longer untethered operation, advancing a fully implantable PSU CVAD toward a critical unmet need in pediatric advanced heart failure.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Multiblade pump: a pLVAD with pump head of high folding ratio for reduced risk of hemolysisJournal of Physics: Conference Series · 2026
- A Novel Extracorporeal Membrane Oxygenation System With Enhanced Hemocompatibility and Long-Term Stability.ASAIO journal (American Society for Artificial Internal Organs : 1992) · 2026
- Fuzzy Logic-Based Modeling and Multi-Objective Optimization of Blood Pump Rotational Speed Considering Hydraulic and Hemolytic Performance.Medical engineering & physics · 2026
- Tolerance-aware robust ranking of ventricular assist device rotor designs using strict feasibility and percentile-based criteria.The International journal of artificial organs · 2026
- Ventriculo-Arterial Coupling and Speed Modulations in Rotodynamic Blood Pumps.American journal of physiology. Heart and circulatory physiology · 2026
- Surrogate-assisted multi-objective optimization of an axial-flow ventricular assist device rotor using a large CFD-derived design space.Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine · 2026