From the Key Laboratory of Biomechanics and Mechanobiology (Beihang University), Ministry of Education; Key Laboratory of Innovation and Transformation of Advanced Medical Devices, Ministry of Industry and Information Technology; National Medical Innovation Platform for Industry-Education Integration in Advanced Medical Devices (Interdiscipline of Medicine and Engineering); School of Biological Science and Medical Engineering, Beihang University, Beijing, China. · National Center for Cardiovascular Disease, Department of Cardiac Surgery, Fuwai Hospital, Beijing, China.

Preclinical

Key points

From the paper's conclusion
  • The ECMO system demonstrated excellent mechanical stability, effective long-term oxygenation, favorable hemodynamic performance, and outstanding hemocompatibility in large-animal models.

Abstract

To optimize and evaluate the hydrodynamic performance and gas exchange efficiency of a newly developed centrifugal blood pump and membrane oxygenator for extracorporeal membrane oxygenation (ECMO) support, blood pump and oxygenator designs were optimized from a hydrodynamic perspective by adopting a five-blade impeller for the pump and a circular flow path with a central inlet and bottom outlet for the oxygenator. Computational fluid dynamics (CFD) was used to evaluate shear distribution. In vitro experiments assessed pressure-flow characteristics. Long-term ECMO support (14 days) was performed in sheep using both venovenous (VV) and venoarterial (VA) modes. Key parameters, including plasma-free hemoglobin and comprehensive blood tests (biochemistry, coagulation, platelet function), were analyzed. Computational fluid dynamics revealed uniform pump flow with minimal high-shear volume and negligible stagnation in the oxygenator, indicating low hemolysis and thrombosis risk. In vitro, the oxygenator showed lower transmembrane pressure than Quadrox-PLS across 1-7 L/min and higher O2 transfer (1-3 L/min). In vivo, circuit flow was stable (2.2 L/min [VV], 2.5 L/min [VA]) and oxygenator ΔP remained 10-12 mm Hg with stable oxygen exchange (155 ml/min [VV], 165 ml/min [VA]), and plasma-free hemoglobin (pfHb) remained below 50 mg/dL. Three non-device-related deaths occurred, and there were no device-related thrombotic or hemorrhagic complications. The ECMO system demonstrated excellent mechanical stability, effective long-term oxygenation, favorable hemodynamic performance, and outstanding hemocompatibility in large-animal models.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers