Multiblade pump: a pLVAD with pump head of high folding ratio for reduced risk of hemolysis

Zhenyan Zhu, Xiaolong Chen, Yanxi Chen, Yichao Tang

Key points

From the paper's conclusion
  • To address this issue, the present study designed and optimized a pLVAD with a size-adjustable pump head, termed the Multiblade Pump (MBP), whose impeller features a high folding ratio.

Abstract

Currently available percutaneous left ventricular assist devices (pLVADs) are constrained by the dimensions of interventional catheters. To achieve adequate blood flow, these devices are forced to operate at markedly high impeller speeds, which in turn elevate the risk of hemolysis. To address this issue, the present study designed and optimized a pLVAD with a size-adjustable pump head, termed the Multiblade Pump (MBP), whose impeller features a high folding ratio. The pump head of the MBP can be implanted into the human body via a catheter in a folded state (outer diameter of 5 mm) and subsequently expanded to its working configuration (impeller outer diameter of 17 mm, folding ratio of 1:3.4). The optimal structural parameters of the MBP were determined as follows: a blade number of 4, an outlet angle of 85°, and an axial distance between different blade layers of 9 mm. When delivering the required human blood flow, the MBP operates at a rotational speed of only approximately 10473 RPM, while its hemolysis index is substantially lower than that of existing pLVADs. In vitro mock circulation experiments confirmed that the MBP could significantly reduce ventricular load, increase aortic pressure and blood flow, and maintain stable performance over 6 hours of continuous operation. The MBP provides a potential new solution for interventional cardiac assistance with low rotational speed and low risk of hemolysis.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Multiblade pump: a pLVAD with pump head of high folding ratio for reduced risk of hemolysis — Science Skim