Clinical and simulated impact of intraocular lens tilt and decentration based on real-world data and optical simulation
Gonzalo Velarde‐Rodríguez, Nicolás Alejandre Alba, Azahara Sánchez‐Lozano, Luis García‐Onrubia, Ignacio Jiménez-Alfaro Morote, Miguel Faria‐Ribeiro
Hospital Universitario Fundación Jiménez Díaz · University of Minho
Key points
From the paper's conclusion- These data were used to reconstruct each eye in optical simulation software and evaluate four scenarios: aligned, decentered, tilted, and combined decentration and tilt.
Abstract
The aim of this study was to determine whether the optical degradation induced by clinically measured intraocular lens (IOL) tilt and decentration, quantified through individualized synthetic eye models, translates into measurable changes in subjective refraction or visual acuity (BCVA). We included 105 eyes implanted with monofocal aspheric IOL and measured their positioning using AS-OCT. These data were used to reconstruct each eye in optical simulation software and evaluate four scenarios: aligned, decentered, tilted, and combined decentration and tilt. For each case, we calculated wavefront aberrations, three objective refractions, and predicted visual acuity from the radial MTF. The average IOL tilt was 5.19 ± 1.28° and decentration was 0.25 ± 0.13 mm. Coma RMS increased from 0.00 μm in the aligned model to 0.26 ± 0.09 μm in the most altered scenario (p < 0.001), and VSX decreased from 0.80 ± 0.11 to 0.35 ± 0.17. Despite this optical degradation, BCVA remained unaffected, with a mean of 0.00 ± 0.05 LogMAR, and no significant correlation was found with tilt or decentration. The best agreement between simulated and clinical refraction was observed when both tilt and decentration were included. These results suggest that moderate IOL misalignment degrades optical quality without compromising visual acuity in monofocal IOLs, but further studies are needed for multifocal designs.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Extended Depth-of-Focus Intraocular Lens Based on Aberration-Neutral Technique.Ophthalmic & physiological optics : the journal of the British College of Ophthalmic Opticians (Optometrists) · 2026
- Cataract Surgery in Eyes with Congenital Iris-Choroidal Coloboma: Quantitative Assessment of Intraocular Lens Alignment and Optical Quality.Journal of clinical medicine · 2026
- Agreement between slit lamp retroillumination pictures and Scheimpflug retroillumination display for toric IOLs orientation.Journal of cataract and refractive surgery · 2026
- Influencing Factors of Positional Stability of Implantable Collamer Lenses in Terms of Rotation, Decentration, and Tilt in Patients With Myopia.Journal of refractive surgery (Thorofare, N.J. : 1995) · 2026
- Risk Factors Associated With ICL Decentration After ICL Surgery.Journal of refractive surgery (Thorofare, N.J. : 1995) · 2026
- Tolerance of residual myopic refractive errors in eyes with enhanced monofocal intraocular lenses with higher-order aspheric optics.Japanese journal of ophthalmology · 2026